2.4

Electrical Connection

WARNING
VOLTAGES IN EXCESS OF 30 VOLTS (RMS) AC OR 50 VOLTS DC, IN CERTAIN
CIRCUMSTANCES, CAN BE LETHAL. CARE MUST BE TAKEN WHEN WORKING ON
LIVE, EXPOSED CONDUCTORS
Power Supply
Single phase
Power Supply Connection
The unit must be connected to the correct electrical power supply as stated, adjacent to the power
connector.
A qualified technician (see page i) must carry out the following procedure.
A power isolation device must be accessible at all times. This device can be considered as
disconnecting the ADTS power cable or building wall isolation switch. The ADTS front panel switch
is not classed as a power isolator.
C
AUTIONS
T
HE SUPPLY MUST PROVIDE CONNECTION TO A PROTECTIVE GROUND TERMINAL
,
AT ALL TIMES
T
HE POWER SUPPLY CABLE AND CONNECTOR MUST BE CORRECTLY RATED FOR THE POWER SUPPLY
European
Colour
Brown
Blue
Green/Yellow
Make sure that the power supply is off before connecting the power cable to the ADTS.
Fuses
The two fuses, located in the holders and mounted on the front panel, protect the unit. The fuses
are connected in the live and neutral supply circuit and are rated at:
• T5AH 250V
K0553 Revision C
110/230 VAC, 50/60Hz
115 VAC, 400Hz
BE CONNECTED TO THE SUPPLY EARTH
US color
Black
White
Green
200 VA MAX
(
GROUND
Function
Live
Neutral
Protective Earth
(Ground)
. T
HE UNIT MUST
).
2-3
,
.