3-16
Max Outlet
Min (Ratio)
Max (Ratio)
Min Inlet Rate
Max Inlet Rate
Min EPR Rate
Max EPR Rate
Select limits
Shows the current setting. To change the select limits setting:
1. Touch the select limits panel.
2. Touch the required select limits radio button. The select limits radio button panel closes and
the select limits panel shows the new selection.
Edit limits
When selected, this function requires you to enter a PIN code. An existing limit set can be edited using
this function. To edit limits:
1. Touch "Edit limits", a numbered keypad is displayed.
2. Enter your PIN number and touch the "tick" icon. The Edit limits panel opens. Touching the "cross"
icon closes the numbered keypad without making any changes.
3. On the Edit limits panel, touch the panel (User 1 to 5) for the limits to be edited. The current limits
are displayed.
4. Touch a limit, a numbered keypad opens allowing you to edit the current limit.
5. Edit the limit and touch the "tick" icon. The parameter is changed and the keypad closes.
Touching the "cross" icon closes the numbered keypad without making any changes.
Create new limits
When selected, this function requires you to enter a PIN code. A new limits set can be created
following the deletion of an existing limits set or by overwriting existing limits. To create new limits:
1. Touch "Create new limits", a numbered keypad is displayed.
2. Enter your PIN number and touch the "tick" icon. The Create new limits panel opens. Touching
the "cross" icon closes the numbered keypad without making any changes.
3. On the Create new limits panel, touch the panel for the custom limits name to be created or
edited (User 1 to 5). The Custom limits name panel and keyboard opens.
4. Touch the "tick" icon. The message "Name Already Exists. Do you want to Overwrite?" is
displayed.
5. Touch "Yes". The previously selected User 1 to 5 limits panel opens. Touching "No" returns you
to the Custom limits name panel.
6. Touching the desired item in the list opens a numbered keypad.
7. Input the new numbers and touch the "tick" icon. The numbered keypad closes and the new
parameter is shown for the selected item. Touching the "cross" icon closes the numbered
keypad without making any changes.
8. If required, repeat this procedure for other parameters.
9. Touch the Back/Return button to return to the ADTS Settings menu.
ADTS542F/552F/553F/554F User Manual
2200 mbar
0.1
10.0
0
1000 mbar/min
0
60 mbar/min
K0553 Revision C