4.2.3 Setting the Alarm Range Error Response

A Range Error can occur when a measured value is within the capacity of the analyzer, but it exceeds the calibration
range of the probe. They can be either
The moisture.IQ indicates Range Errors with an
measurements of the same mode. For example, if a dew point measurement is
Over Range.
also be
If multiple Range Errors occur simultaneously, the moisture.IQ responds to them in the following
Oxygen Errors
Moisture Errors
order:
,
Complete the following steps to configure the range error response of the selected alarm:
Channel (1-6)
1. Tap the
Figure 49 below.
2. Set the alarm to either
setting).
3. Set the alarm to either
Save
4. Tap the
button to keep the new settings or tap the
moisture.IQ User's Guide
Under Range
or
Over Range
Temperature Errors
,
Alarm
A
B
and
(
or
), and then tap the
Figure 49: Range Error Setup Screen - Alarm Response
Trip
on Under Range errors or to
Trip
on Over Range errors or to
Chapter 4. Using the Outputs, Alarms and Logger Menus
Over Range
errors.
Under Range
or
message.The error condition applies to all
Over Range
Pressure Errors.
and then
Range Error
Ignore
Under Range errors (
Ignore
Over Range errors (
Cancel
button to keep the previous settings.
, then moisture in ppMv will
button to open a screen similar to
Ignore
is the default
Ignore
is the default setting).