9.8 Input Specifications

9.8.1 Capability
Note: Each module can be configured as one or three channels.
1 to 6 channels moisture
1 to 6 channels temperature
1 to 6 channels pressure
1 to 6 channels oxygen
2 auxiliary inputs per installed channel
Note: Inputs may be used for 0 to 2 mA, 4 to 20-mA, and voltage-based transmitter in the -1 to +4-V range including
devices such as Oxygen Analyzers, Thermal Conductivity Analyzers, Flowmeters, Pressure Transmitters,
Temperature Transmitters, etc.
9.8.2 Resolution
16 bits
9.8.3 Moisture Sensor Probes
GE types: M-Series, Moisture Image Series, and MISP2.
9.8.4 Temperature Sensor
Thermistor (optionally supplied as part of the moisture probe assembly).
9.8.5 Pressure Transmitter
A pressure transducer is optionally available for Moisture Image Series probes. GE P40, P40X, or equivalent 4-20 mA,
current-transmitting, pressure transducer; scale factors are entered as part of the user-program sequence.
9.8.6 Intrinsic Safety
Built-in intrinsic safety provided for all inputs per drawing #752-364, excluding auxiliary inputs.

9.9 Probe Specifications

9.9.1 Moisture Image Series Probe or MISP2
9.9.1a
Type
Aluminum oxide moisture sensor probe and electronics module
9.9.1b
Calibration
Each sensor is individually computer calibrated against known moisture concentrations,
traceable to national standards.