CALIBRATION

Temperature calibration

1. Press the CAL key for approximately 5 seconds.
2. The third digit of the temperature indicator flashes, and the CO
3. Set the present correct temperature with the
4. The unit automatically reverts to the display mode.
[Example]
If the displayed chamber temperature is 37.0
1. Press the CAL key for about 5 seconds.
2. The "3" on the temperature indicator flashes, and the CO
3. Adjust the set value to the actual value of 36.8
4. The unit automatically reverts to the display mode.
Note:
It is important to accurately measure the temperature inside the unit when performing temperature
calibration. Particularly, the temperature gauge used must have an accuracy of 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting must not change by more than ±1.0
error tone is emitted, the input data is ignored, and the unit reverts to the display mode. Consequently, if
it is necessary to change the temperature by more than 1.0
a period of time.
density indicator goes out.
2
key and
key, then press the ENT key.
o
C (set value) and the actual temperature is 36.8
density indicator goes out.
2
o
C with the
key and
o
C during calibration. If it exceeds this, an
o
C, perform calibration in several stages over
92
o
C.
key, then press ENT key.
36