CALIBRATION
Temperature calibration
1. Press the calibration key (CAL) for approximately 5 seconds again to enter the temperature calibration
input mode.
2. The left digit of the digital temperature indicator blinks, and the digital CO
density indicator goes out.
2
3. Set the present correct temperature with the digit shift key (
) and numerical value shift key (
) ,
then press the enter key (ENT).
4. The incubator automatically reverts to the display mode.
[Example]
o
o
If the displayed chamber temperature is 37.0
C (set value) and the actual temperature is 36.8
C.
1. Press the calibration key (CAL) for about 5 seconds.
2. The "3" on the digital temperature indicator blinks, and the digital CO
density indicator goes out.
2
o
3. Adjust the set value to the actual value of 36.8
C with the digit shift key (
) and numerical value shift
key (
), then press enter key (ENT).
4. The incubator automatically reverts to the display mode.
31