BEFORE COMMENCING OPERATION
Connection of gas injection nozzle
In the case of control of O
density in the chamber, connect the gas injection nozzle to the gas injection
2
port by using the connecting pipe enclosed (inner diameter; 5mm, outer diameter; 9mm, length; 300mm).
This helps faster recovery of humidity after opening of the inner door.
Fill the humidifying pan with sterile distilled water so that the injection nozzle can be under the water
level.
1. Inside of unit is arranged as shown in Fig.1 when humidifying pan cover removed. There is a gas
injection port in the lower left-hand corner.
2. Connect gas injection nozzle to gas injection port using connecting pipe. (Fig.2)
3. Connecting gas injection nozzle is as shown in Fig.3.
Connecting pipe
Gas injection nozzle
Gas injection port
Fig.1
Fig.2
Fig.3