ALARMS & SAFETY FUNCTIONS
Alarm & Safety
If the ambient temperature sensor is
Ambient
disconnected.
temperature sensor
abnormality
If the ambient temperature sensor is
short circuited.
CO
sensor
If the output voltage of the CO
2
abnormality
is abnormal.
O
sensor
If the output voltage of the O
2
abnormality
is abnormal.
If the upper limit alarm temperature
Main heater
alarm operates, or if the main heater is
abnormality
open circuit, or the main heater relay
short circuit.
Bottom heater
If the bottom heater goes open circuit,
abnormality
or the bottom heater relay short circuit.
Door heater
If the door heater goes open circuit,
abnormality
or the door heater relay short circuit.
Disconnection of
If the relay of main heater, bottom
S.S.R for each
heater or sensor box heater goes open
heater
circuit.
Low humidifying
If the water in the pan is about 0.8 liter. RH PAN lamp blinks.
water
UV lamp failure
When the UV lamp is burned out.
The accumulated ON time is over
Recommendation of
about
new UV lamp
1,000 hrs.
• The alarm can be canceled by pressing the BUZZER key, but the remote alarm cannot be silenced. And the upper
limit temperature alarm cannot be silenced with the BUZZER key.
• E01 or E02 is cleared automatically when the gas is connected correctly and the buzzer is silenced with the
BUZZER key. Press BUZZER key to silence the alarm after changeover of gas supply line.
• If one of E05 to E17 is displayed, consult with a Sanyo sales representative or agent.

Operation after power failure

The set value is memorized by nonvolatile memory. Accordingly, the incubator resumes the operation
with setting before power failure.
Situation
E09 is displayed alternately
with the temperature on the
temperature indicator.
E10 is displayed alternately
with the temperature on the
temperature indicator.
E11 is displayed alternately
sensor
2
with the temperature on the
temperature indicator.
E19 is displayed alternately
sensor
2
with the temperature on the
temperature indicator.
E12 is displayed alternately
with the temperature on the
temperature indicator.
E13 is displayed alternately
with the temperature on the
temperature indicator.
E14 is displayed alternately
with the temperature on the
temperature indicator.
E16 is displayed alternately
with the temperature on the
temperature indicator.
E18 is displayed alternately
with the temperature on the
temperature indicator.
UV indicator blinks when
UV lamp is OFF.
86
Indication
Buzzer
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
-----
Intermittent tone
-----
Safety operation
Remote alarm
CO
valve closes.
2
Remote alarm
N
/O
valve closes.
2
2
Remote alarm
Remote alarm
Remote alarm
Remote alarm
Remote alarm
-----
Remote alarm
-----
30