Installation Of H Generator (Mco-Hp) - Sanyo MCO-19M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

Installation of H
Prior to use H
O
Generator, ensure to check the following notes.
2
2
After installation
Make sure the condensing fan should be correctly installed to the position.
Make sure the duct should be correctly mounted to the position.
Make sure the H
O
2
2
each fixture should be correctly mounted to the holes at the bottom of the duct.
"Err34" (The volume of H
If you find an error message, 'Err34' after predetermined decontamination process is
complete, the voltage of the apparatus may be out of the specification and it is necessary to
adjust H
O
generating time. Please consult to your authorized sales representative.
2
2
* Secure H
O
Generator with the fixtures to prevent generator from being moved or rattled.
2
2
O
Generator (MCO-HP)
2
2
Generator should be fixed by 2 fixtures which are pointed in the picture and
O
generated is abnormal)
2
2
Fixture left and right
- 87 -
Duct
H
O
Generator (MCO-HP)
2
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19m(uv)