Sanyo MCO-19M Service Manual page 30

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

3. MENU/Set - Various settings
<Setting of chamber temperature>
In Top Screen, press MENU key and select Set.
Press ENTER key to display current setting temperature in the right side of "Temperature " column
in "Stand-by Setting" display.
Input required value and press MENU key and select OK in reverse video.
Press ENTER key to memorize the input value.
Setting range is 0 ~ 50.0 .
Note) Maximum 50.0
<Setting of CO2 density>
In Top Screen, press MENU key and select Set.
Press ENTER key to display current setting CO2 density in the right side of "CO2 Density" column
In "Stand-by Setting" display.
Input required value and press MENU key and select OK in reverse video.
Press ENTER key to memorize the input value.
Setting range is 0 ~ 20.0%.
Note) Maximum 20.0% for settable value. Error tone is not emit if you input a value over 20.0%.
<Setting of O2 density>
In Top Screen, press MENU key and select Set.
Press ENTER key to display current setting O2 density in the right side of "O2 Density" column
In "Stand-by Setting" display.
Input required value and press MENU key and select OK in reverse video.
Press ENTER key to memorize the input value.
Setting range is 1~80.0%. (Controllable range: 1~18.0%, 22.0%~80.0%)
Note) O2 is set to 1.0% when you input a value lower than 1.0% and is set to 80.0% when you input a value
over 80.0%. Error tone is not emit.
Note) Press ENTER key with "Cancel" displayed in reverse video to cancel the setting change.
<Display of high limit setting temperature>
High limit setting temperature is displayed in the right side of "High Limit" column in "Stand-by Setting" display.
Setting range is 35 ~51 .
When a chamber temperature is over high limit setting temperature, LED illuminates and audible alarm sounds
continuously. (The audible alarm will not stop sounding if BUZZER key is pressed)
for settable value. Error tone is not emit if you input a value over 50.0 .