16 H2O2 decontamination
<Setting of H2O2 decontamination>
Ensure first that UV lamp function and H2O2 decontamination mode are active.
Select "System Configuration" in "Select Tools Svc" display.
Press MENU key and select "OK".
Press ENTER key to display "System Configuration".
Ensure "1" is displayed both in the right side of "UV Function" column and "Decon Function" column.
Note) When "0" is selected in both "UV Function" column and "Decon Function" column, H2O2 decontamination
mode is inactive,if H2O2 key is pressed for 3 seconds.