Sanyo MCO-19M Service Manual page 82

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

(2) H
O
generator (MCO-HP), H
2
2
Installation
Fill 80cc distilled water in MCO-HP. By using the cable enclosed, connect MCO-HP with the connector
on the bottom right of the incubator. (See picture as below)
Connector cap should be kept in safe place.
Ensure 2 pins of MCO-HP should be aligned to the holes on the rear duct. If not so, it may be caused
poor in spray performance.
For the check of spray performance, hold door switch by the tape.
Front
H
O
generator
2
2
Operation check
#
1
2
Press H
O
key for 3 seconds.
2
2
3
4
Ensure if the outer door is locked by
electric key.
5
Open outer door by using key to check
spray performance is correctly obtained.
O
Decontamination kit (MCO-HL)
2
2
Rear duct
UV lamp
Connector
Ultrasound oscillator
Operation
H
O
generator
2
2
Key
-----
H2O2
H2O2 Decontamination screen
'H2O2 Decon' blinks
-----
(If error message is shown, refer
to Instruction Manual P.57.)
-----
-----
Turn key toward the direction
when you locked and take key off
before door shut. You cannot turn
the key with the door shut.
- 79 -
UV lamp
Front
Display/Remark
"Top Screen"
"H2O2 Decontamination"
(System check starts)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19m(uv)

Table of Contents