<Error message in H2O2 decontamination>
(Note Only 1 error message is displayed even if multiple error messages are present.
Error #
Status
Err 31
When
H2O2
disconnected
Err 32
When H2O2 solution has not
been filled or H2O2 sensor
fails (disconnected)
Err 33
When a door is open
Err 34
When H2O2 is abnormally
generated
(If H2O2 misting time is over
15minutes than predetermined
time)
(May be caused by faulty
oscillator, faulty fan, poor fan
operation or H2O2 generator
was set incorrectly)
Err 35
When power fails during H2O2
is generated
Err 36
When UV lamp fails
Err 37
When power fails during UV
resolution process
generator
is
"H2O2 Decon System Error"
"Err31: H2O2 Unit No Connect"
"H2O2 Decon System Error"
"Err32: H2O2 Level is Low"
"H2O2 Decon System Error"
"Err32: H2O2 Level is Low"
"Err34: H2O2 Volume"
"Err35: Power Failed" is displayed
after power returns.
"Decon Failed" is displayed after UV
resolution process is complete.
"Err36: UV Lamp Failer"
"Err37: Power Failed" is displayed
after power returns
"Decon Finished" is displayed after
UV resolution process is complete.
Display
- 69 -
Alarm
Action
Decontamination
mode aborts
Decontamination
mode aborts
Decontamination
mode aborts
Buzzer
UV resolution will
sounds
be commenced
intermittently
Remote
alarm
activates
Buzzer
UV resolution will
sounds
be
commenced
intermittently
after
returns
Remote
alarm
activates
Buzzer
UV
resolution
sounds
time
will
intermittently
extended
Remote
alarm
activates
UV resolution will
be
commenced
again after power
returns
power
be