Procedure; Table 10-7 Procedure For Entering Values For Lead Resistance Compensation - Honeywell 9782 Series Operator's Manual

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Procedure

To use the 9782's automatic compensation for lead resistance, enter the necessary values using
the procedure in Table 10-7.

Table 10-7 Procedure for Entering Values for Lead Resistance Compensation

Step
1
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV |
MAINTENANCE
OFFLINE FUNCTIONS
INSTRUMENT SETUP
LEAD RESISTANCE COMP
DIAGNOSTICS
| PREV |
2
LEAD RESISTANCE COMP
WIRE GAUGE
WIRE LENGTH
CELL1
CELL2
DIAGNOSTICS
| PREV |
7/99
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
Screen
Go to MAINTENANCE | LEAD RESISTANCE COMP.
NEXT
NEXT
Use this screen to specify the gauge used for leads to both
cells, and the length of one lead to each cell.
12AWG
100FT
If mixed gauges are used, or the length exceeds 1999 feet,
150FT
refer to the Introduction of this sub-section for guidance in
NEXT
computing the values to enter here.
Maintenance
Action
10-13