Conductivity/Resistivity Analyzer/Controller
Step
AUTO CLEAN
3
TIMER SETUP
DAYS
HOURS
MINUTES
CURRENT DAY
DIAGNOSTICS
|
AUTO CLEAN
4
RINSE SETUP
DURATION
RESUME DELAY
DIAGNOSTICS
| PREV |
5-20
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
Screen
Timer setup is used to specify the frequency of the rinses.
3
16
30
2
| NEXT
using the real time AutoClean Time Remaining display:
Specify where today falls in the AutoClean cycle. For
example, if the interval for rinses is 3 days and you want
the first rinse to be tomorrow, specify that today is day 2.
Enter the duration of the rinse (in seconds).
Enter the number of seconds the Analyzer/Controller
180
60
should wait after the rinse before resuming normal
operation.
NEXT
If several rinses are to be performed in a single day:
a) specify the number of days as zero
b) specify the number of hours and minutes in the
interval between rinses.
If rinses are to occur at intervals of days:
a) specify the number of days in the interval
b) specify the time of day (hours and minutes) the
rinse should occur.
In our example, the rinse will occur every third day at
4:30 PM.
If you plan to use AutoClean, but initiate rinses manually
a) specify the number of days as zero
b) specify the number of hours and minutes in the
interval between rinses as zero also.
IF YOU ENTER ALL ZEROS, NO AUTOMATIC
RINSES WILL BE PERFORMED.
Action
7/99