9782 Series Two Cell Conductivity/Resistivity Analyzer/Controller Operator’s Manual 70-82-25-74 Rev 3 6/99...
Page 2
In no event is Honeywell liable to anyone for any indirect, special or consequential damages. The information and specifications in this document are subject to change without notice.
7/98 12/98 6/99 References Honeywell Documents The following list identifies all Honeywell documents that may be sources of reference for the material discussed in this publication. Document Title Trademarks AutoClean is a trademark of Honeywell Inc. Noryl is a trademark of GE Company.
All 9782 models can use relays for control...1-16 Optional analog outputs for control ...1-16 Selecting a strategy ...1-16 1.6 Overview of Installation and Setup Tasks... 1-17 Setup tasks described in this manual...1-17 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Contents 6/99...
Page 5
3. UNPACKING, PREPARATION, AND MOUNTING ... 3-1 3.1 Overview ... 3-1 Introduction ...3-1 What’s in this section? ...3-1 3.2 Unpacking and Preparing ... 3-2 Procedure...3-2 3.3 Mounting ... 3-3 Introduction ...3-3 Panel-mounting...3-3 Wall- and pipe-mounting ...3-3 6/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual...
Page 7
7.1 Overview ... 7-1 Introduction ...7-1 What’s in this section? ...7-1 7.2 AutoClean Sequence and Piping ... 7-2 Rinse sequence...7-2 Select valves and fittings with appropriate pressure ratings...7-3 Minimize liquid volume in system...7-3 6/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual...
Page 8
Self-tests at 0.5 second intervals ...9-4 Output affected by error...9-4 9.5 Off-Line Diagnostics ... 9-5 Introduction ...9-5 9.5.1 Display and Keyboard Test ... 9-5 Procedure...9-5 9.5.2 Output Tests ... 9-7 Procedure...9-7 viii 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 6/99...
Page 9
What’s in this section? ...11-1 11.2 Part Numbers ... 11-2 Introduction ...11-2 APPENDIX A PROPORTIONAL CONTROL TUNING ... A-1 Adjust effective proportional band width ...A-1 Treatment takes too long, but no unacceptable overshoot ...A-1 Unacceptable overshoot...A-1 6/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual...
Page 10
SW2 - settings are dependent on options ... C-1 APPENDIX D USING A PRECISION CHECK RESISTOR...D-1 Introduction ...D-1 Set cal factor and calibration trim for ideal conditions ...D-1 Calculations for conductivity, resistivity, and TDS ...D-1 Concentration values...D-2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 6/99...
Page 11
Table 10-4 Procedure for Adjusting the Screen Contrast ... 10-9 Table 10-5 Procedure for Entering a Password... 10-10 Table 10-6 Procedure for Resetting All Configuration to Factory Settings... 10-11 Table 10-7 Procedure for Entering Values for Lead Resistance Compensation... 10-14 6/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual...
Page 12
Table 10-8 Procedure for Replacing the Fuse... 10-16 Table 11-1 Part Numbers ... 11-2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 6/99...
Page 13
Figure 6-3 Interconnections between 9782 and Recorder with Range Identification by Point Change 6-11 Figure 7-1 Automatic Cell Wash Setup ... 7-3 Figure 10-1 Example of a Conductivity Loop... 10-14 Figure 10-2 Location of Power Fuse... 10-18 6/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual xiii...
Page 14
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 6/99...
1.1 Description Multi-function instrument The Honeywell 9782 Conductivity/Resistivity Analyzer/Controller (Figure 1-1) is a microprocessor-based instrument for analysis of conductivity, resistivity, total dissolved solids, or concentration in industrial processes. The resolution for all measurements is +/- 0.5% of reading (except for temperature; see below). The units of measure are:...
Conductivity/Resistivity Analyzer/Controller Input Input to the 9782 can be from any Honeywell conductivity cell at distances up to 1500 feet. Outputs Optional isolated 0 to 1 V, 0 to 10 V and 4 to 20 mA outputs are available. Use these analog outputs for retransmission of process variables, or for Current Adjusting Type control using an output signal that is directly proportional to the input (see 1.5.3).
If a problem is detected, a message is displayed on the alarm stripe to alert the operator. In addition, the operator can initiate keypad and display tests using Maintenance Menu functions. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Introduction...
For power plant morphonline conductivity measurements temperature compensation was developed specifically for this market’s water chemistry. Morpholine has unique temperature characteristics that are handled automatically by the 9782 Analyzer/Controller. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
Readout in conductivity or ppm TDS units may also be selected. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Figure 1-1 9782 Front Panel Introduction...
DOWN ENTER F1, F2, F3 [function keys] 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Table 1-1 Key Functions When process values are on display: Use DISPLAY to cycle through available real-time displays. When the main menu (Configuration, Calibration, Maintenance, I/O Setup) is on display: Use DISPLAY to return to displaying process values.
AutoClean was scheduled or started manually. The alarms and outputs will be held throughout the AutoClean operation. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual soft keys to move the cursor to the digit to be changed Introduction...
(without going through the menus), and to acknowledge alarm stripe messages. Figure 1-2 Representative Process Value Display for a Cell in Standard Range Analyzer 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual CONDENSATE T105 8.59...
AutoClean is active, the screen will display the time remaining. When this screen is on display, the function keys can be used to control the AutoClean operation as described in 1.3.1. 1-10 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual AUTO CLEAN TIME REMAINING...
CELL CAL FACTOR CALIBRATION TRIM TDS CONVERSION FACTOR C1 TDS CONVERSION FACTOR C2 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Purpose set up Analyzer/Controller to implement your alarm and control strategy configure parameters for using analog outputs for control and retransmission of...
Page 26
SCREEN CONTRAST SECURITY RESET UNIT LEAD RESISTANCE COMP [wide range models only] I/O SETUP 1-12 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Purpose do occasional system tasks [see functions listed below] test display test keyboard test relays and analog outputs...
1. Cell 1 “Advanced Features” AutoClean and range status indication 2. discrete control (both cells) 3. alarm annunciation (both cells), and Cell 2 AutoClean and range status 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Introduction 1-13...
“or discrete control” will not be available in that row’s combination of relay assignments. If the Analyzer/Controller has only two relays, then only the choices for Relay 1 and Relay 2 will be available. 1-14 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
If you decide to use proportional-only control, select the type based on the input requirements of the device to be controlled. See Appendix A for a discussion of tuning proportional-only control. 1-16 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
Display model number stored in memory and software version number. Note these for future reference. (If you call for technical assistance, you will need this information.) 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Task Introduction Section 1-17...
Page 32
(Section 10) instructions for calibrating the outputs and changing the fuse (Section 10) parts list (Section 11) supplementary information for special applications and proportional control tuning tips (appendices). 1-18 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
3-2 or the Parts List. 2.1.3 Compliance FM/CSA This product has FM and CSA Hazardous Location Approval for Class I, Div. 2, Groups A, B, C and D classified areas. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Specifications and Model Number...
9782C- _ _ - C2: 9782C- _ _ - C3: One analog output also available in combination with optional relays (see above). 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual RH; 120 or 240 Vac; process temp 25 C per C; RH:...
DAT cycle period: 1 to 200 seconds. PFT maximum frequency: 1 to 200 pulses/minute. PFT pulse width: 50 ms, compatible with electronic pulse-type metering pumps. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual acid/cation/ammonia, or morpholine. NaOH, or H resistivity: 0 to 19.99 Mohms/cm...
Key Number _ _ _ _ _ KEY NUMBER pH/ORP Conductivity Specific Ion 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual (No other options from Table II) _ _ _ _ _ _ Description Specifications and Model Number Selection Availability...
Page 40
Factory Set 120 Volt Operation Factory Set 240 Volt Operation Certificate of Calibration TABLE IV Factory Use Only 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Standard Wide (Relay contacts rated for 120 volt maximum) (Note 2) (Note 3) (Note 4)
Page 41
20-1000 mS Weight % Concentration 0-18.00% Hydrochloric Acid 0-19.99% Sulfuric Acid 0-19.99% Sodium Hydroxide 0-19.99% Sodium Chloride 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Specifications and Model Number STANDARD RANGE Cell Constant 0-1999 ppb 0.01 0-19,990 ppb 0-199.9 ppm...
Section 5. What’s in this section? The topics in this section are listed below. 3.1 Overview 3.2 Unpacking and Preparing 3.3 Mounting 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Unpacking, Preparation, and Mounting Topic See Page...
Carefully remove the instrument from the shipping container. Compare the contents of the shipping container with the packing list. Notify the carrier and Honeywell immediately if there is equipment damage or shortage. Do not return goods without contacting Honeywell in advance.
ATTENTION Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration may affect system performance. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Honeywell York, PA U.S.A. ATTENTION - UNIT SET FOR 240 VAC Directions: 70-82-25-74...
The topics in this section are listed below. 4.1 Overview 4.2 General Wiring Practices for Power Wiring 4.3 Power Wiring Considerations 4.4 Installing Power Wiring 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 4. Power Wiring Topic Power Wiring See Page...
Figure 4-1. The wound filter cores shall be located external to the 9782 case within 25 cm. Conform to code Instrument wiring should conform to regulations of the National Electrical Code. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
Power supply voltage and frequency within specs The power supply voltage and frequency must be within the limits stated in the specifications in Section 2. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Power Wiring ) wire for ac power.
Failure to observe this precaution may result in serious personal injury. Replace the safety cover and secure it with the screw. Close the case and power up the unit. Do not apply power until the case is closed. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action 7/99...
Power Wiring Power Terminations L2/N PE = Protective Earth Ground Figure 4-1 AC Power Terminals 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual...
I/O setup if the measurement type, relay activation, display resolution, cell constant, or an analog or discrete I/O assignment is changed. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual I/O Setup and System Configuration Topic See Page...
CLOCK SETUP: A display showing the current time can be accessed when the unit is online. The 24-hour clock must be set before you can use the AutoClean feature described below. Set the clock as described in 5.7. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
Table 5-10 contains the menu choices available (or ranges for numeric values) and system defaults for I/O assignments and configuration parameters, as well as for values entered during maintenance functions described in Section 10. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual I/O Setup and System Configuration...
For both Cell 1 and Cell 2, use AutoClean (Relays 1 and 3), with no range status indication, nor any discrete control. For both Cell 1 and Cell 2, use range status indication (Relays 2 and 4), with no AutoClean, nor any discrete control. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
Page 61
As a result, you will be able to configure the unit to take advantage of features requiring that hardware. Procedure The procedure for I/O setup configuration is shown in Table 5-1. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual I/O Setup and System Configuration...
If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE ENABLED”, instead of displaying output assignments. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to I/O SETUP. The first screen shows the present use of available relays.
Page 63
DIAGNOSTICS | PREV | NEXT 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual If any relays are still available, the screen will display a query about discrete control. If discrete control will be used, select the relay(s) to be used and the type of control for each.
Page 64
The cell calibration factor is found on the plastic tag hanging from the cell lead wires. It is entered using the calibration menu as described in Section 8. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Use this screen to specify the display resolution, LOW (fixed) or HIGH (automatically adjusted).
Page 65
This output range scaling has no effect on the display. Display resolution is configured using the maintenance menu as described in Section 10. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Select the type of solution for which temperature compensation should be applied.
If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE ENABLED”, instead of displaying analog outputs. 5-10 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Process Variables Action Go to CONFIGURATION | CAT/RETRANSMISSION.
Page 67
DIAGNOSTICS | PREV | NEXT 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Select an output to configure it for control. Go to the next screen. The input type and appropriate unit of measure will already be filled in, based on the process variable or computed value assigned to this analog output during I/O setup.
The procedure for configuring On/Off control is shown in Table 5-3. The procedure for configuring PFT control is shown in Table 5-4. The procedure for configuring DAT control is shown in Table 5-5. 5-12 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Cell Cell Cell...
250 µS ONTIME DIAGNOSTICS 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to CONFIGURATION | DISCRETE CONTROL | ON/OFF. ATTENTION: Any value entered for one of these control- related parameters will take effect as soon as it is entered.
PB LIMIT 250 µS DIAGNOSTICS 5-14 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to CONFIGURATION | DISCRETE CONTROL | PFT. ATTENTION: Any value entered for one of these control- related parameters will take effect as soon as it is entered.
PB LIMIT 250 µS DIAGNOSTICS 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to CONFIGURATION | DISCRETE CONTROL | DAT. ATTENTION: Any value entered for one of these control- related parameters will take effect as soon as it is entered.
DEADBAND 250 µS DIAGNOSTICS 5-16 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to CONFIGURATION | ALARMS. ATTENTION: Any value entered for one of these alarm- related parameters will take effect as soon as it is entered. Use this screen to enter a delay time to prevent brief process upsets from activating an alarm.
HOURS MINUTES DIAGNOSTICS NEXT 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Table 5-7 Procedure for Setting the Clock Go to CONFIGURATION | ADVANCED FEATURES | CLOCK SETUP. Set the clock. For example, if you will start the clock at 1:30 PM, enter 13 hours, 30 minutes.
Once the AutoClean sequence has been completed (including resume delay), the “AUTO CLEAN” message disappears and the time remaining clock will be displayed again, showing the time remaining until the next scheduled timer-initiated cleaning. 5-18 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
BE SETUP FOR THIS FEATURE DIAGNOSTICS | NEXT 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual I/O Setup and System Configuration Action Go to CONFIGURATION | ADVANCED FEATURES | AUTO CLEAN. If you have already set the clock, go to the next screen.
Page 76
| PREV | NEXT 5-20 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Timer setup is used to specify the frequency of the rinses. If several rinses are to be performed in a single day: a) specify the number of days as zero b) specify the number of hours and minutes in the interval between rinses.
NEXT ADVANCED FEATURES CLOCK SETUP COMPUTED VALUE AUTO CLEAN DIAGNOSTICS | PREV | NEXT 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual I/O Setup and System Configuration Cell Cell Cell Cell Cell Cell Action Go to CONFIGURATION | ADVANCED FEATURES | COMPUTED VALUE.
Page 78
Conductivity/Resistivity Analyzer/Controller Step Screen Action COMPUTED VALUE Specify the value to be computed. SELECT DIFFERENCE DIAGNOSTICS NEXT 5-22 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
I/O setup: cell constant for Cell 1 I/O setup: cell constant for Cell 2 I/O setup: AutoClean enabled I/O setup: range status enabled 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual I/O Setup and System Configuration CHOICES or RANGE (Default is underlined) 9782C-S0...
Page 80
I/O setup: display unit of measure (conductivity only) I/O setup: relay alarm assignments I/O setup: analog output basis [unless DE is enabled] configuration: retransmission output scaling 5-24 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual CHOICES or RANGE (Default is underlined) O/F [on/off] ENERGIZE...
Page 81
AutoClean timer set day configuration: AutoClean timer set hour configuration: AutoClean timer set minutes 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual I/O Setup and System Configuration CHOICES or RANGE (Default is underlined) range depends on measurement type and cell constant;...
Page 82
TDS conversion factor 5-26 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual CHOICES or RANGE (Default is underlined) 0 to 28, default = 0 1 to 1999 seconds, default = 60...
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open the case while the unit is powered. Disconnect power before installing I/O wiring. More than one switch may be required to disconnect power. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Topic Input and Output Wiring See Page...
In applications where either the power, input or output wiring are subject to electromagnetic disturbances, shielding techniques will be required. Grounded metal conduit with conductive conduit fittings is recommended. Conform to code Instrument wiring should conform to national and local electrical codes. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
Cables for locations without interference For steel (metal) conduit or open wiring away from any sources of interference, use recommended Honeywell cables listed in the parts list in Section 11. Shielded wiring for locations with interference In applications where plastic conduit or open wire trays are used, shielded 6-conductor 22 gage (0.326 mm...
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open the case while the unit is powered. Disconnect power before installing I/O wiring. More than one switch may be required to disconnect power. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Cell Cell...
If the relays will not be used, replace the safety cover and secure it with the screw. Close the case and power up the unit. Do not apply power with the case open. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Input and Output Wiring Action...
If a relay is being used to indicate automatic output range change, see 6.4.3 for guidance in wiring a device to the relay. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Table 6-3 Relay Contact Maximum Ratings 0.6 amps at 120/240 Vdc, 0.6 amps at 110 Vdc,...
Replace the safety cover and secure it with the screw. Close the case and power up the unit. Do not apply power until the case is closed. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Table 6-4 Procedure for Wiring Relays...
Replace the safety cover and secure it with the screw. Close the case and power up the unit. Do not apply power until the case is closed. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Input and Output Wiring Action...
The resistance value can be anything from 500 ohms to 1 Mohm. Range 1 Range 2 Figure 6-2 Sample Multipoint Recorder Chart Illustrating Auto Output Range Change for Steam Condensate Monitoring 6-10 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual (simulated condenser leak) a/n 23178 7/99...
9782 Conductivity Analyzer Figure 6-3 Interconnections between 9782 and Recorder with Range Identification 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Multi-point Recorder Input Terminal Board by Point Change Input and Output Wiring where R = 250 ohms a/n 23179...
Install piping and valves as diagrammed in this section. What’s in this section? The topics in this section are listed below. 7.1 Overview 7.2 AutoClean Sequence and Piping 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual AutoClean Theory and Piping Topic See Page...
Note that the operator can make the operation pause using the “soft key” on the special AutoClean Time Remaining display. If the operator does not remove the pause by pressing the PAUSE soft key again, the Analyzer/Controller will resume normal operation after 20 minutes. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
Minimize liquid volume in system Keep pipe sizes small and couplings close to minimize the liquid volume in the system. Smaller volumes require less time to rinse. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual rinse/cleaning solution 9782 cell...
Each type of cell has an associated cell constant entered during I/O setup (see Section 5). This number is part of the cell model number. However, for greater precision, every Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead wires.
Page 100
If both cells’ solution temperature compensation is set up with TDS, then the calibration menu will be: Therefore, the screens on your unit may not match exactly the calibration menus used in the procedure tables in this section. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual CELL CAL FACTOR CAL TRIM C1 CAL TIRM C2...
If the cell constant is 0.01, 0.1, or 1.0, run a wire from the black electrode terminal of the cell to the lower card retainer bracket screw. Do not ground cells having cell constants of 10, 25, or 50. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Calibration...
The constant for each cell is entered during I/O setup. However, for greater precision, every Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead wires.
De-ionized water to rinse the cell. Table 8-2 Conductivity of Potassium Chloride Solutions at 25 C Concentration M* * M = Molarity; 1M = 74.555g potassium chloride per liter of solution 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Conductivity (microsiemens per cm) 0.001 0.005 0.01...
147.00 µS DIAGNOSTICS CAL TRIM C1 147.00 µS CAL ADJ FAILED 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Prepare container of the reference solution. Go to CALIBRATION | CAL TRIM C 1. Remove the cell from the process. Rinse the cell thoroughly with de-ionized water.
DIAGNOSTICS REMOVE CAL TRIM DIAGNOSTICS 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to CALIBRATION | REMOVE CAL TRIM C1. A calibration trim message will be displayed. Go to the next screen. To specify that you want to remove the trim for the cell, change the NO to YES.
For example, if you plan to use “NaCl (TDS)” set the solution temperature compensation type to “NaCl” temporarily for calibration purposes. (Solution temperature compensation type is the one I/O setup parameter that can be changed without triggering a cold reset.) 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
REMOVE CAL TRIM C1 REMOVE CAL TRIM C2 DIAGNOSTICS 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to CALIBRATION | TDS CVSN FACTOR C1. If you plan to calibrate, do it before entering the theoretically derived TDS conversion factor (see previous page).
The topics in this section are listed below. 9.1 Overview 9.2 System Status Messages 9.3 Process Alarm Messages 9.4 On-Line Diagnostics and System Error Messages 9.5 Off-Line Diagnostics 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Diagnostics and Messages Topic See Page...
RETRIEVING DATA ENTERING DATA INPUT RANGE TOO HIGH INPUT RANGE TOO LOW 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Table 9-1 System Status Message WHEN DISPLAYED This message is displayed temporarily and indicates that an AutoClean operation is active. All alarms and outputs are held for the duration of the AutoClean operation.
TEMP 2 HIGH ALM TEMP 1 OR 2 HIGH ALM TEMP 1 OR 2 LOW ALM 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Table 9-2 Process Alarm Messages WHEN DISPLAYED Indicates alarm state, if the Analyzer/Controller is configured to alarm on the Cell 1 and/or Cell 2 value exceeding (high alarm) or falling below (low alarm) the configured alarm setpoint.
Note 1: Temperature compensator should be 8550 ohms at 25 C. If temperature compensator fails, a conductivity instrument will be forced to its display range high limit. A resistivity instrument will be forced to the low end of the range or zero. 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual WHEN CONDITION...
DIAGNOSTICS DTEST| DISPLAY TEST F1=TOGGLE MENU=ABORT DIAGNOSTICS 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Diagnostics and Messages Action Go to MAINTENANCE | OFF-LINE FUNCTIONS | DISPLAY TEST. Pressing the specified function key will display a test pattern on the first line of the display. Press the key repeatedly to test the rest of the display.
Page 114
KEYBOARD TEST PRESS A KEY TO TEST MENU TO EXIT DIAGNOSTICS 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Go to MAINTENANCE | OFF-LINE FUNCTIONS | KEYBOARD TEST. Press any key (except MENU) to display the name of the key.
Table 9-5 Procedure for Testing Relays and Analog Outputs Step Screen 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Turn off the power to the Analyzer/Controller. More than one switch may be required to remove the power. With the power off open the case:...
Page 116
A A N A A N N N DIAGNOSTICS | PREV | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Go to MAINTENANCE | OFF-LINE FUNCTIONS | OUTPUT/RELAY TEST. Select the combination of output values to be supplied to the terminals, or go to the next window for additional choices.
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open the case while the unit is powered. More than one disconnect switch may be required to remove power. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 10. Maintenance Topic Maintenance...
Page 118
The steps indicated above must be followed to prevent damage and/or degradation, which may be induced by ESD, to static sensitive devices. 10-2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
A disconnect switch must be installed to break all current carrying conductors. Turn off power before working on conductors. Failure to observe this precaution may result in serious personal injury. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Maintenance 10-3...
KEYBOARD TEST OUTPUT/RELAY TEST OUTPUT CALIBRATION DIAGNOSTICS 10-4 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Turn off the power to the Analyzer/Controller. More than one switch may be required to disconnect power. With the power off open the case:...
OUTPUT CALIBRATION SAVED MENU TO EXIT DIAGNOSTICS 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action The first display will reflect the quantity of analog outputs supported by the hardware. Select the value to be calibrated. Go to the next screen.
UPDATE TAG DATA HONEYWELL INC DIAGNOSTICS | <-- | --> 10-6 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Go to MAINTENANCE | INSTRUMENT SETUP | INSTRUMENT TAGGING. The text string currently stored in memory will be displayed. If desired, edit the string (nineteen characters maximum) using the UP and DOWN keys, and the <-- and -->...
A special password is required to change the stored model number. If the password was not supplied with the new option card, call the Honeywell Technical Assistance Center. Before calling for technical support, follow the procedure in Table 10-4 to determine the version of the software in the Analyzer/Controller.
Page 124
DIAGNOSTICS If you do not have the factory password, call the Technical Assistance Center at 1-800-423-9883. 10-8 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual The software version is read-only information. A1.0 If the model number displayed no longer matches the...
USE UP/DOWN ARROW KEYS TO ADJUST ENTER TO SAVE DIAGNOSTICS 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Go to MAINTENANCE | INSTRUMENT SETUP | SCREEN CONTRAST. Use the UP and DOWN keys to adjust the contrast between 0 and 20.
PASSWORD SECURITY ENABLE SECURITY: YES PASSWORD: DIAGNOSTICS 10-10 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to MAINTENANCE | INSTRUMENT SETUP | SECURITY. To enable the security function, select YES. The current password will be displayed. It can be edited 0000 using this screen.
WILL BE RESET TO FACTORY DEFAULTS! DIAGNOSTICS RESET| 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Go to MAINTENANCE | INSTRUMENT SETUP | RESET UNIT. Pressing the specified function key at this point will perform a cold reset as described above.
Only) Introduction If you use standard Honeywell cell lead lengths of 7 or 20 feet connected directly to the Analyzer/Controller, no compensation for lead resistance is necessary. Similarly, if a junction box is used to extend the leads up to 150 feet, no compensation is required. However, if longer leads are used (greater than 150 feet), signal quality can be adversely affected unless you enter information that will permit the 9782 to compensate for lead resistance.
| PREV | NEXT 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Go to MAINTENANCE | LEAD RESISTANCE COMP. Use this screen to specify the gauge used for leads to both cells, and the length of one lead to each cell.
To avoid potentially serious damage to expensive components, a blown fuse should not be replaced until the cause has been determined by a qualified service technician. Procedure To replace the fuse follow the procedure in Table 10-8. 10-14 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Figure 10-2 Location of Power Fuse 7/99...
Page 131
If any leads were disconnected in Step 6, reconnect them. Replace the safety cover and secure it with the screw. Close the case and power up the unit. Do not apply power until the case is closed. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Action Maintenance 10-15...
This section provides part numbers for field-replaceable parts and for accessories. What’s in this section? The topics in this section are listed below. 11.1 Overview 11.2 Part Numbers 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Accessories and Replacement Parts List Topic See Page 11-1 11-2...
Pipe/wall mounting kit 084711 Panel mounting kit 056348 056349 Reducer panel kit 11-2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Table 11-1 Part Numbers Description Parts Included in Kit upper card retainer bracket lower card retainer bracket screws for retainer brackets...
Page 135
31316260 aluminum junction box for use with additional cable for distances beyond 150 ft Note 1: Order cable by the length in whole foot increments. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Description AVAILABLE CABLES Description Description installation instructions...
Reduce the reagent delivery capacity. Reduce the reagent concentration. Widen the proportional band by adjusting the proportional band limit further from the setpoint. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix A - 1 Appendix A Proportional Control Tuning...
Page 138
Conductivity/Resistivity Analyzer/Controller Appendix A - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/99...
F ee d Cell Cell Percent Passage = Cell Cell Percent Rejection = 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix B - 1 Appendix B Cell 2 Cell 2 Cell 2 Cell 1 Cell 1 Anion Bed C e ll 1...
Cell 2 Rinse Water Cell Conductivity Ratio = Cell Appendix B - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 9782 Analyzer Cell 2 Cell 1 Cell 1 Parts Flow Typical Ratio Range is 0.1 to 1.1 Output Signal...
9782 includes reboiler measurement functions to automate ASTM Test Method D4519 for determining anions and carbon dioxide. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix B - 3 Cell 2 Cell 1 (Soft) Treated Water Typical Ratio Range is 1 to 1.25...
Operation above these limits gives a flashing display. The measurement range of hydrochloric acid is restricted to less than 15.5% above 40º C and less than 18% below 40º C. Appendix B - 4 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Specific Conductivity...
The settings of the eight switches in SW2 depend on the options used. The appropriate settings are shown in Table C-1. 7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix C - 1 Microprocessor Board Switch Settings Appendix C Switch 1...
Page 144
Table C-1 Microprocessor Board Switch Settings Option None Single Parameter Current or Voltage Output Auxiliary Relays Two Analog Outputs Three Analog Outputs Appendix C - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual SW2-1 SW2-2 SW2-3 SW2-4 SW2-5 SW2-6 SW2-7 SW2-8 7/99...
Analyzer/Controller input terminals. In addition, an 8550 ohm resistor (Honeywell Part No. 233300) can be wired in place of the inputs from the temperature compensator to simulate 25º C, the reference temperature. The unit will display a simulated “process value”...