Honeywell 9782 Series Operator's Manual page 116

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Conductivity/Resistivity Analyzer/Controller
Step
Screen
7
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV | NEXT
MAINTENANCE
OFFLINE FUNCTIONS
INSTRUMENT SETUP
LEAD RESISTANCE COMP
DIAGNOSTICS
| PREV |
OFF-LINE FUNCTIONS
DISPLAY TEST
KEYBOARD TEST
OUTPUT/RELAY TEST
OUTPUT CALIBRATON
DIAGNOSTICS
| PREV |
OUTPUT TEST
01
02
03
0
0
25
0
50
25
75
50
DIAGNOSTICS
|
9
OUTPUT TEST
01
02
03
100
75
0 100
0
0 100
0
0
DIAGNOSTICS
| PREV |
10
11
12
13
9-8
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
Go to MAINTENANCE | OFF-LINE FUNCTIONS |
OUTPUT/RELAY TEST.
NEXT
NEXT
Select the combination of output values to be supplied to the
R1R2R3R4
terminals, or go to the next window for additional choices.
0
A A A A
0
A A A A
When one of the rows of output combinations has been
0
A A A A
25
A A N A
selected, observe the meter readings to check that the analog
|
NEXT
output values match those selected. Also listen for the "click"
of an energized relay.
If the results of the analog output test are not satisfactory,
recalibrate the output's zero and span using the procedure in
10.2.
If the results of the relay output test are not satisfactory,
consult Honeywell for assistance.
Select one of these combinations, or use the specified function
R1R2R3R4
key to return to the previous display.
50
A A A N
75
A N A A
A A N A
0
A N N N
If only one meter is available, so that only one output can be
tested at a time, then for each analog output repeat Steps 1
through 9, including powering down the unit before
changing the connections to the output terminals.
When output testing has been completed, re-install field wiring
removed in Step 4. Disconnect power before opening case.
Replace the safety cover and secure it with the screw.
Close the case and power up the unit. Do not apply power
until case is closed.
Action