Motion Limit Checks
4.5.4.1

Motion Controller Software 7MOT0096-N14 and Later

CAUTION
The ranges within which the ultimate switches must go active are:
Horizontal:
Vertical:
Rotational:
Note. It is not possible to approach the rotational ultimate limit by rotating the detector to the Z
< – 90° limit. However, testing the ultimate limit at Z
of the functionality, since the same limit switch serves both rotational limits.
Preparing for Testing Ultimate Limit Switches
Testing ultimate limit switches requires the following preparations:
Removing both table console covers (front and rear – if these have not already been removed)
Disabling software limits. This is done by sending commands to the master (rotate) motion
controller via a HyperTerminal connection
Starting the CardioMD acquisition software.
Before testing X axis limits, furthermore:
Removing the patient table pallet.
Before testing Y axis limits, furthermore:
Removing the patient table pallet
Rotating the detector assembly to approximately 110°.
Disabling Software Limits
In order to verify ultimate switches, software limits must first be disabled (the diagram in Figure
4.1 on page 4-40 may give you an idea, why). In order to accomplish this, software limits are tem-
porarily set to values well beyond the ultimate limits. For this, you need to establish a Hyper-
Terminal connection to the rotation motion controller (the motion controller to the right).
For instructions on using HyperTerminal for communicating with motion controllers, see Chap-
ter 5 Diagnostics, the section Motion Controller Diagnostics. Once the HyperTerminal connection
is established:
Procedure
1. When you see the prompt
HyperTerminal window indicating that
you are communicating with the motion
controller, enter the following string of
commands:
wrsdo 0x607D 1 -752400 [Enter]
wrsdo 0x607D 2 3762000 [Enter]
wrsdo 0x687D 1 -59904 [Enter]
wrsdo 0x687D 2 8815104 [Enter]
wrsdo 0x707D 1 -5184000 [Enter]
wrsdo 0x707D 2 5184000 [Enter]
4-42 CardioMD Service Manual
Caution. With motion controller software release 7MOT0096-N14 and later,
SW8 on the motion controllers can no longer be used.
– 7 mm < X
< – 2 mm, 497 mm < X
min
218 mm < Y
< 228 mm, 1195 mm < Y
min
Z
> 182°.
max
Details
in the
->
< 505 mm
max
< 1199 mm
max
> 182° alone is sufficient for verification
max
1MAN0158-H11 04-2005 CSIP Level 1