Detector Calibration Files

Procedure
> CardioMD [Enter].
The acquisition PC answers: Welcome to
Microsoft Telnet Server. You can now enter
DOS commands to operate the acquisition PC.

5.10 Detector Calibration Files

The directory C:/Cardiocam/Calibrations stores five calibration files per detector. The
files are in ASCII format and can be reviewed in Windows Notepad. The five files are:
dyncor.det#
Ecal.det#
Geocal.det#
HVcal.dat
PMTcal.det#
where
is
#
Figure 5.16 on page 5-59 shows typical values in the files together with a brief explanation.
5-58 CardioMD Service Manual
for detector 1 and
for detector 2.
1
2
Details
1MAN0158-H11 04-2005 CSIP Level 1