Collimator Id Programming; Tools Required; Procedure - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Gantry Software Updates

Procedure

11. When both EDC modules have been up-
dated, remount the detector top cover fol-
lowing instructions provided in Chapter 6.
Repair Procedures.
7.2.4

Collimator ID Programming

The collimator type ID is programmed into an EEPROM located on the collimator contact board.
7.2.4.1

Tools Required

CardioMD software installation CD, part number 9CSY0906
NULL modem serial interface cable (DB9-female to DB9-female).
When using a laptop acquisition PC (CardioMD Series III): An USB to RS 232 adapter, part
III
no. 3ACQ1669.
7.2.4.2
Procedure
Before starting the collimator ID programming, check the release note on the CD-ROM. You find
the release note in the directory
is more recent than this manual.
Procedure
1. Remove the detector top cover to access
the EDC modules. See instructions in
Chapter 6 Repair Procedures.
2. Remove the ECG cable from the ECG port
labeled A1 RS232.
3. Connect one end of the serial interface
cable to this port.
If you are using a laptop acquisition PC:
4. Use the USB to RS 232 adapter to connect
the opposite end of the serial interface ca-
ble to one of the acquisition PC's USB
ports.
See section 7.2.2, page 7-2 for instructions on
determining which COM port you are using.
If you are using a tower acquisition PC:
5. Connect the opposite end of the cable to
the acquisition PC's COM1 serial port.
7-12 CardioMD Service Manual
Details
. The release note may provide information that
Release notes
Details
III
I, II
Serial interface
cable
ECG cable
1MAN0158-H11 04-2005 CSIP Level 1