Procedure
When both the X and Y measurements have
been completed, a pop-up similar to the one
shown on the right appears.
14. Note the results and click OK to close the
pop-up.
Resolution specifications are:
≤
UFOV FWHM
≤
CFOV FWHM
≤
UFOV FWTM
≤
CFOV FWTM
Linearity specifications are:
≤
UFOV absolute
≤
CFOV absolute
UFOV differential ≤
CFOV differential ≤
15. Replace the NEMA Y line phantom with
the collimator.
16. Click Exit to return to the NEMA menu.
Note. To measure the other detector, place the CardioMD system in the patient load position be-
fore repeating the procedure outlined above.
1MAN0158-H11 04-2005 CSIP Level 1
Details
3.7 mm
3.7 mm
7.6 mm
7.6 mm
0.5 mm
0.5 mm
0.2 mm
0.2 mm.
NEMA Performance Measurement
CardioMD Service Manual 10-11