Saving Site Data

1-52
The currently programmed site data may be stored in the Model
DF868's non-volatile memory by saving it as a site file. Up to ten site
file names, consisting of up to five characters each, may be stored at
any given time. While following the instructions, refer to Figure A-5
in Appendix A, Menu Maps. Enter the
[F4]
and
at the initial User Program screen and complete the
following steps:
1. Enter a new Site File Name and press
[F1]-[F4]
and
keys to select and overwrite an existing site file.
Note: The option bar will show only as many choices as there are
existing site files.
If ten site files are already stored in memory, an existing site file must
be deleted before a new site file name can be assigned. See Chapter 5,
Clearing Data, for details.
When a site file is stored in memory, it becomes the active site file
until another site file is created or retrieved. (See the next section for
instructions on retrieving a site file.) To verify that the site file has
been saved, use the [↑] key to scroll back to the
and see if the name appears on the option bar (use the [←] and [→]
keys to scroll through the list, if necessary). Whenever a new site file
name is saved, it is automatically assigned to the next available
[EXIT]
function key. Press
2. Use the [←] and [→] and
[EXIT]
press
to leave the User Program to return to measurement
mode.
SAVE
submenu by pressing [→]
[ENT]
or use the [←], [→]
SITE NAME
SAVE
to leave the
submenu.
[F1]-[F4]
keys to select another menu, or
Programming Site Data
prompt