Bep2 Cover And Octopus Card Holder Replacement Procedure - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-5-3

BEP2 Cover and Octopus Card Holder Replacement Procedure

NOTE:
BEP2 refers to a Pentium 4-based BEP (newer type).
8-5-3-1
Tools
Use the appropriate flat and Phillips-type screw drivers, as indicated in the BEP cover replacement
procedure.
8-5-3-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
8-5-3-3
BEP2 - Side Cover Removal Procedure
1) Remove the two screws (that secure the side cover in position) from the rear of the BEP (see
8-59
2) Slide the cover backwards and remove it.
8-5-3-4
BEP2 - Side Cover Installation Procedure
1) Return the cover to the side of the BEP and slide it forwards into position.
2) Replace and tighten the two screws to hold the cover firmly in position
BEP2" on page
8-64
04
EVISION
below).
Screw
Screw
Figure 8-59 Side Cover - BEP2
8-64).
Section 8-5 - Back End Parts Replacement
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
Chapter 3 -
Installation.
(Figure 8-59 "Side Cover -
M
ANUAL
Figure

Advertisement