Input Ac Voltage Configuration; Secondary Voltage Configuration; Ac Input Cord - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
Section 6-2

Input AC Voltage Configuration

The internal Vivid 3N voltage is 220V AC.
The Back End and Front End Crates operate on 220V AC.
The Monitor and Peripherals operate on local voltage.
6-2-1

Secondary Voltage Configuration

The AC input voltage is factory preset according to specific requirements, as indicated on the label near
the AC input connector. The AC transformer's primary and secondary voltages are set up by using the
appropriate plugs connected in Jumper1 and Jumper2 on the AC Distribution Box, in accordance with
the information in
Table 6-2 Primary and Secondary AC Voltages
Jumper 1
PRIMARY
Jumper 2
SECONDARY
6-2-2

AC Input Cord

The input voltage power cord and plug must meet the local standards and requirements.
6-2
04
EVISION
Table 6-2
below:
Voltage
Range
Plug P/N
Voltage
Plug P/N
Section 6-2 - Input AC Voltage Configuration
100-120V
220-240V
FK200077
FK200078
100V
120V
FK200079
FK200080
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
220-230V
FK200084
M
ANUAL
230-240V
FK200085