Signal Conditioning, Self Calibrating Digital Indicators
Title
Error
Error 32
BOGUS EE ADDR
Error 33
SIG.MOD WIRES
Error 34
SIG.MOD MISMATCH
Error 35
BOGUS MENU TYPE
Error 36
BAD EE TABLE
Error 37
BAD PERSONALITY
Error 38
BAD FORMAT
Error 39
WRONG HARDWARE
Error 41
SHUNT CAL ERROR
Error 43
BAD IDENT CRC
Error 44
BAD FUNCTION
Error 45
A/D NAK
Error 46
SENSE WIRES
Error 47
BAD EE TYPE REQ
Error 48
NULL MENU
Error 49
NULL ACTION FCN
Error 50
EXCITATION
Error 51
CAN'T DO EXC.
Error 52
SIG.MOD LENGTH R
Error 53
SIG.MOD RECTYPE
74
Honeywell Sensing and Control
Description
An invalid address has been passed to the programmer. Restart the instrument. If this
problem persists, contact our Service department with the information found in the SYSTEM
MENU -> SOFTWARE REV. menu item.
The instrument has detected that the (+)Signature and (-)Signature pins (labeled as
"+MEM" and "-MEM) are likely reversed.
The excitation type byte found inside a Signature Module indicates that the channel is not
designed to operate with that type of transducer.
A menu contains an invalid menu type. Contact Honeywell with the information found in
the SYSTEM MENU -> SOFTWARE REV. menu item.
The instrument doesn't understand the EEPROM type setting for the channel. Contact
Honeywell with the information found in the SYSTEM MENU -> SOFTWARE REV. menu item.
The channel doesn't understand the given card type. Contact Honeywell with the informa-
tion found in the SYSTEM MENU -> SOFTWARE REV. menu item.
The channel's EEPROM memory contains a display setting which is not understood by
the instrument. Check that every item in the channel's DISPLAY SETUP menu has a valid
setting.
The channel type expected by the instrument for the channel number is not the channel
type that is physically installed. If you have changed the address jumpers on one or more
channels, replace them in their original configuration.
The application of the shunt resistor did not affect the transducer's output. Check the wir-
ing and the transducer output with a voltmeter to verify proper operation. If this is a Strain
Gage Channel, make certain that the shunt calibration resistor installed on the circuit
board is the correct value per the transducer's Calibration Record. If this is a High-Level
Input Channel, make certain that the Shunt Cal 1 and Shunt Cal 2 pins are connected to
the correct pins of the transducer/amplifier.
A checksum failure has occurred when reading a channel's EEPROM. Restart the instru-
ment.
The instrument has requested that the channel perform a function that it is not capable of.
Restart the instrument.
Hardware error. The channel's Analog-to-Digital converter is not communicating with the
channel's microprocessor. Restart the instrument.
The Strain Gage Channel's (+)Sense and (-)Sense pins (labeled as "+SEN" and "-SEN")
are not connected to the (+)Excitation and (-)Excitation pins (labeled as "+EXC" and
"-EXC"), respectively, This is most likely caused by the connector being unplugged from
the channel. If you are not interested in the using this channel and you wish to suppress
this error, see "DISABLE CHANNEL Menu Item" on page 36.
When installing a channel, the EEPROM type requested by the channel is not valid. Con-
tact Honeywell with the information found in the SYSTEM MENU -> SOFTWARE REV. menu item.
A menu contains a NULL item. Contact our Service department with the information found
in the SYSTEM MENU -> SOFTWARE REV. menu item.
A menu contains a NULL action function. Contact our Service department with the infor-
mation found in the SYSTEM MENU -> SOFTWARE REV. menu item.
The channel's excitation voltage regulator has shut down because the protection circuit
that limits the output current has activated. This is most likely caused by a short of the (+)
Excitation and (-)Excitation pins (labeled as "+EXC" and "-EXC").
The channel's EEPROM memory contains an excitation setting which is not understood
by the instrument. Change the channel's CALIBRATION DATA -> EXCITATION menu item to
another setting.
An unexpected length byte was found inside a Signature Module during a read operation.
Check the wiring.
An unexpected record type was found inside a Signature Module during a read opera-
tion. Check the wiring.