Auxiliary Output Printed Circuit Board Kit
(Part Number 51404561-501 or 51404561-503)
Summary
Kit Contents
Form 44-45-33-52A
DR 4500A Circular Chart Recorder
Kit Instruction
Form: 44-45-33-52A
Effective: 1/15
Supercedes: 5/98
Enclosed with this instruction is a new Auxiliary Output printed circuit
board for your DR4500A Recorder (Classic or Truline models).
Use the procedures listed in the tables to install, configure, and calibrate
the Auxiliary Output Printed Circuit board. These procedures assume that
you are familiar with the DR4500A recorder. We suggest that you have a
copy of the recorder's product manual on hand for reference.
ATTENTION
Kit Part Number 51404561-503 contains a printed circuit
board that has both Auxiliary output and RS485 Modbus RTU functions.
Refer to Instruction Sheet 44-45-33-58 for information on the RS485
Modbus RTU function.
WARNING
Never access components inside the case with the power
applied.
CAUTION
When handling electronic assemblies, care should be taken to
avoid damage by electrostatic discharge. Use a wrist strap connected to an
earth ground before withdrawing the printed circuit board from its
container and installing it into the recorder.
This kit contains parts to add Auxiliary Output to your DR4500A recorder.
You kit you have received contains the following parts:
Part Number
51404561-001
or
51404561-003
30756005-001
30755139-002
30366609-004
NOTE: You may wish to change or add Table III in your recorder's model
number to equal 1XX to show that the recorder is equipped with the
Auxiliary Output function, or 4XX to show that the recorder is equipped
with the Auxiliary Output and Modbus RTU Communication function.
Auxiliary Output Kit Instruction
Description
Auxiliary Output PCB
or
Auxiliary Output/Modbus RTU Communications PCB
Card Guide (2)
Terminal Block - 2 Position
Tie Wraps (2)
Continued on next page
Page 1 of 8

Advertisement

loading

Summary of Contents for Honeywell DR 4500A

  • Page 1 DR 4500A Circular Chart Recorder Kit Instruction Auxiliary Output Printed Circuit Board Kit (Part Number 51404561-501 or 51404561-503) Form: 44-45-33-52A Effective: 1/15 Supercedes: 5/98 Summary Enclosed with this instruction is a new Auxiliary Output printed circuit board for your DR4500A Recorder (Classic or Truline models).
  • Page 2 Auxiliary Output Installation Installation Procedure Follow the steps listed in Table 1 to add the 4-20mA Auxiliary Output printed circuit board. The procedure assumes that the chart door is open, the chart plate is swung out, and the power is removed. Table 1 Adding a 4-20mA Auxiliary Output PCB Step...
  • Page 3 Auxiliary Output Installation, Continued Installation Procedure, Figure 1 shows the location of the Auxiliary Output components. continued Figure 1 Auxiliary Output Components Auxiliary Output Modbus Motherboard Modbus Connector Transformer Connector Controller Load 0 - 1000 Ohms – Plug J2 Auxiliary Output Connector 22849C Continued on next page...
  • Page 4 Auxiliary Output Configuration Enable auxiliary output The procedure to enable Auxiliary Output is listed in Table 2. Table 2 Procedure for Enabling Auxiliary Output Step Operation Press Action/Result Access the Until you see: Options Set Upper Display Up group SET UP Lower Display OPTIONS Access the...
  • Page 5 Auxiliary Output Configuration, Continued Parameter selections Table 3 lists the available Auxiliary Output Selections and their definitions. Table 3 Auxiliary Output Selection and Definitions Lower Display Upper Display Parameter Prompt Range of Setting Definition or Selection AUX OUT AUXILIARY OUTPUT SELECTION provides an mA output representing any of twelve control parameters.
  • Page 6 Auxiliary Output Configuration, Continued Parameter selections, continued Table 3 Auxiliary Output Selection and Definitionscontinued Lower Display Upper Display Parameter Prompt Range of Setting Definition or Selection AUX OUT OUT 2 OUTPUT 2—Represents the displayed controller output in percent (%) for Loop 2. (continued) SP 2 SETPOINT 2—Represents the value of the setpoint in...
  • Page 7: Auxiliary Output Calibration

    Auxiliary Output Calibration Introduction Calibrate the recorder so that the auxiliary output provides the proper amount of current over the desired range. • The recorder can provide an auxiliary output current range of from 0 to 21 milliamperes and can be calibrated at 4 mA for 0% of output and 20 mA for 100% of output or any other values between 0 and 21 Equipment needed You will need a standard shop type milliammeter, with whatever accuracy...
  • Page 8 Auxiliary Output Calibration, Continued Procedure The procedure for calibrating the Auxiliary Output is listed in Table 5. Make sure “LOCKOUT” in the Set Up group is set to “NONE.” Also, “AUX. OUT” must NOT be disabled in the “Options” Set Up group.11 Table 5 Auxiliary Output Calibration Procedure Step...