TITLE
CN70 Service Work Instructions
Repair & Test Flow
OPERATION:
The following repair and test flow process must be followed to ensure that a unit being repaired is properly tested and
configured.
CN70 high Level repair process
Step Process
1
Repair Unit
Review Service
2
Repair Bulletins
3
USB PreFlash
70-Series BIT Test
4
Procedure
CN70 OS Flash
5
Update Procedure
Load OS per unit's
5
configuration
Load SU per unit's
6
configuration
Government
Configurations
7
Load Information
Assurance Software
Reference
Document
Document
Location
SWICN70-000
WorkManager
Service Repair
CK70 SRI
Bulletins
page links
OTPCN70-006
WorkManager
OTPCN70-009
WorkManager
OTPCN70-005
WorkManager
SAP BOM
SAP
SAP BOM
SAP
943-418-002
WorkManager
Model
DOCUMENT NO.
CN70A,CN70E
SWICN70-000
Comments
Erase FFS, load Default CStore, load Non-
LP OS
Run built in test procedure (installs and
runs on the unit being tested)
Alternate process used in manufacturing to
load the core operating system. Not
typically used in service.
OS installed using an SD card. This step is
key on LP (Language Provision) OS
configurations.
Load SU (Software Update). An SU of
zero zero doesn't need to be reloaded.
Link to software download can be found
on the CN70 SRI page
PAGE
Page 3 of 45