2.3 Operating Page

2.3.1 Single Input pH

7. 0 0
pH
24. 4
Deg.C
Monitoring pH
404
mV
Millivolts
100. 0
%
7. 0 0
pH
A: Slope & Check
6
Measured Values
pH.
Temperature.
Measured Millivolts
Millivolts.
% Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see
Section 5.3, page 30) and 105% is displayed. If the value is outside these limits, check
the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7 pH is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
See Section 3.1, page 12.
VIEW SETPOINTS
See Section 4.1, page 19.
SENSOR CAL.
set to
A3: Type
Wash Function
not set to
A3: Type
Monitoring pH
(Section 5.5) – see Section 2.3.6, page 11.
Wash
(Section 5.5) – return to top of page.
Wash
2 Operation
– see
Set Min Slope
IM/AX4PH Rev. P