Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
6.6.2 Differential ph Systems Connections – Capable of Providing Sensor Diagnostics (AP200, TBX5)
When connecting one of these pH systems to the AX400 transmitter, ensure the differential input switch for the relevant sensor is
See Section 5.4, page 33 for sensor diagnostics configuration. If diagnostics are not required leave them switched off.
Terminal Block B
Sensor B
Sensor A
B1
B9
B2
B10
B3
B11
B4
B12
B5
B13
B6
B14
B7
B15
B8
B16
Table 6.3 Differential ph Systems Connection – AP200, TBX5
*In normal operation do not connect the heavy green wire. If noisy readings are obtained, connect the wire to the earth stud.
Note.
1. If the sensor is fitted with a 2-wire PT100, Pt1000 or 3K Balco temperature compensator, link terminals B9 and B10 (and
B1 and B2 if dual input analyzer.).
2. Redox systems are not temperature compensated so do not have temperature sensors. To remove temperature error
messages, set
Temperature Sensor
to the correct type – see Section 5.3, page 30.
Temperature Sensor
IM/AX4PH Rev. P
Function
Temperature Compensator (if fitted) Common –
see also Note 1below
Temperature Compensator (if fitted) 3
Temperature Compensator (if fitted)
Reference
Not Used
Solution Earth (Ground Rod)
Screen/Shield (if fitted)
Glass/Metal Electrode
to
. If a temperature sensor is used for a separate temperature display, set
NONE
AP200
Color
Grey
rd
Lead
White
Green
Blue
Not Used
Green/Yellow
Red
Clear
6 Installation
.
ON
*TBX5
Color
White
Link to White
Red
Black
Not Used
Green
Yellow
Blue
55