5.3 Configure Sensors

-----
CONFIG. SENSORS
-----
Config. Sensor A
ORP
Redox
-----
pH
A: Probe Type
Yes
-----
No
A: Diff. Input
A: Probe Type
Redox/ORP
None
3K Bal
Pt1000
-----
Pt100
A: Temp. Sensor
Yes
-----
No
A: Enable Cals
Config. Sensor A
30
Configure Sensor A
Sensor B configuration (dual input analyzers only) is identical to Sensor A
Config. Sensor B
configuration.
Probe Type
Select the required probe type.
ORP
Redox
pH
Differential Input
Yes
No
Note. Ensure the electrode system is connected correctly for the type of input
selected (Standard or Differential) – see Fig. 6.9 (wall-/pipe-mount analyzers) or
Fig. 6.11 (panel-mount analyzers).
set to
Probe Type
A: Temp. Sensor
set to
Probe Type
A: Electrode
Temperature Sensor
Select the type of temperature sensor used: Pt100, Pt1000, Balco 3K or None.
Enable Calibration
Select
to enable sensor calibration.
Yes
Select
to disable sensor calibration and all associated frames.
No
Return to main menu.
CONFIG. SENSORS
See Section 5.4, page 33.
CONFIG. ALARMS
Millivolt display
Millivolt display
pH display
Select if electrode system is equipped with a solution earth (ground)
rod and sensor diagnostics are required – see Section 5.4, page 33.
Select if electrode system is not equipped with a solution earth
(ground) rod or sensor diagnostics are not required.
or
– continued below.
Redox
ORP
– continued on next page.
pH
5 Programming
IM/AX4PH Rev. P