Re-Calibration – 5 Point START
To begin the 5-point re-calibration procedure, turn the unit OFF
and follow the steps below.
Keystroke
1. Press and hold the
MIN/MAX key and turn
the unit on by pressing
the ON/OFF key.
2. Press the arrow key
until "START" is
displayed on the bottom
line.
3. Press the PRGM key.
4. Press the arrow key
until "5-POINT" is
displayed on the bottom
line.
5. Press the PRGM key.
6.
Vent P1 and P2 ports to
atmosphere and
simultaneously press the
MIN/MAX and HOLD
keys, then release.
7. Press the arrow key
while on the right most
digit to proceed.
8. Apply the indicated
calibration point pressure
using external pressure
standards. After pressure
is stable, press the
arrow key.
9. Repeat step 8 for CAL
POINTS 4 and 5.
30 of 36
Display
Top line reads "RE-CAL".
Bottom line reads "EDIT".
Top line reads "RE-CAL".
Bottom line reads "START".
Top line reads "RE-CAL
Bottom line reads "1-
POINT".
Top line reads "RE-CAL
START".
Bottom line reads "5-
POINT".
Top line reads "POINT 1 –
ZERO:"
Bottom line displays live
applied pressure.
Unit takes new zero. Top line
reads " POINT 1 - ZERO:"
Bottom line displays live
applied pressure. POINT 1
has been taken.
Top line reads " POINT 2 -
APPLY:".
Bottom line displays the cal
point value to apply.
Top line reads " POINT 3 -
APPLY:".
Bottom line displays the cal
point value to apply.
Top line reads "POINT 4/5 -
APPLY"
Bottom line displays the cal
point value.