Re-Calibration – 5 Point START
To begin the 5-point re-calibration procedure, turn the unit OFF
and follow the steps below.
Keystroke
1. Press and hold the
MIN/MAX key and turn
the unit on by pressing
the ON/OFF key.
2. Press the  arrow key
until "START" is
displayed on the bottom
line.
3. Press the PRGM key.
4. Press the  arrow key
until "5-POINT" is
displayed on the bottom
line.
5. Press the PRGM key.
6.
Vent P1 and P2 ports to
atmosphere and
simultaneously press the
MIN/MAX and HOLD
keys, then release.
7. Press the  arrow key
while on the right most
digit to proceed.
8. Apply the indicated
calibration point pressure
using external pressure
standards. After pressure
is stable, press the 
arrow key.
9. Repeat step 8 for CAL
POINTS 4 and 5.
30 of 36
Display
Top line reads "RE-CAL".
Bottom line reads "EDIT".
Top line reads "RE-CAL".
Bottom line reads "START".
Top line reads "RE-CAL
Bottom line reads "1-
POINT".
Top line reads "RE-CAL
START".
Bottom line reads "5-
POINT".
Top line reads "POINT 1 –
ZERO:"
Bottom line displays live
applied pressure.
Unit takes new zero. Top line
reads " POINT 1 - ZERO:"
Bottom line displays live
applied pressure. POINT 1
has been taken.
Top line reads " POINT 2 -
APPLY:".
Bottom line displays the cal
point value to apply.
Top line reads " POINT 3 -
APPLY:".
Bottom line displays the cal
point value to apply.
Top line reads "POINT 4/5 -
APPLY"
Bottom line displays the cal
point value.