Re-Calibration - 5 Point Edit - Omega HHP360 Series User Manual

Precision absolute manometer
Table of Contents

Advertisement

Re-Calibration – 5 Point EDIT
To edit the calibration points for a 5 Point re-calibration follow the
steps below.
NOTE: If the factory default values are acceptable, skip this
section and proceed to the re-calibration 5-Point START procedure.
Keystroke
1. With unit OFF, press and
hold the MIN/MAX key,
turn the unit on using the
ON/OFF key, then release
2. Press the PRGM key.
3. Press the  or  arrow
keys to edit the selected
digit. Use the  or 
arrow keys to change the
cursor position.
Note: For 0% go directly
to step 4.
4. Press the  arrow key
while on the right most
digit to proceed.
5. Repeat steps 3 and 4 for
CAL POINTS 2, 3, 4 and
5.
6. After editing CAL
POINT 5 press the 
arrow key while on the
right most digit to
proceed.
7. To perform the 5-point re-
cal, press the  arrow
key until START is
displayed on the bottom
line.
OR
To exit without
performing the 5-point re-
cal press the  arrow key
29 of 36
Display
Top line reads "RE-CAL".
Bottom line reads "EDIT".
Top line reads "CAL POINT
1". Bottom line displays the
cal point value.
Bottom line displays the cal
point value. The cursor
flashes at the first position
while the value is changed,
then moves to the right
position when the right arrow
key is pressed.
Top line reads "CAL POINT
2". Bottom line displays the
cal point value.
Top line reads "CAL POINT
2/3/4/5". Bottom line
displays the cal point value.
Top line reads "RE-CAL".
Bottom line reads "EDIT".
Top line reads "RE-CAL".
Bottom line, "START".
Continue with 5-Point Re-
calibration procedure at step 3
on next page.
OR
Returns to Measure Mode.