Lockout Select

Enabling the Lockout feature prevents unauthorized users from
making changes to the configuration of the manometer. To enter
the Program Mode, the user must first enter the "password" (two-
digit Lockout Code) within approximately 40 seconds when
prompted.
Failure to enter the correct two digit code within
approximately 40 seconds will return the unit to Measure Mode.
Any two-digit numeric code can be programmed. The factory
Lockout Code of 00 (which is the default as shipped from the
factory) disables the Lockout.
To set the Lockout Code follow these steps:
Keystroke
1. From the Measure Mode
press the PRGM key. If
the Lockout is set, enter
the correct "password"
when prompted.
2. Press the  arrow key
twice.
3. Press the  arrow key
then the  arrow key
four times.
4. Press the  arrow key,
then press the  or 
arrow keys to change the
first digit. Press the 
arrow key to proceed.
5. Press the  arrow key
when the desired code is
set. Lockout is activated.
6. Press the  arrow key
twice.
22 of 36
Display
Top line reads "PROGRAM
MODE" and bottom line
reads "UNITS SELECT".
Bottom line reads "USER
INFO SELECT".
Top line reads "LOCKOUT
CODE" and bottom line reads
"ENTER TO SELECT".
Bottom line shows the old
Lockout Code. The cursor
flashes at the first position
while the value is changed,
the cursor moves to the right
position once the right arrow
key is pressed.
Top line reads "LOCKOUT
CODE" and bottom line reads
"ENTER TO SELECT".
Returns to Measure Mode.