■ Calibration

This function corrects the deviation of the measurement value caused from
the color, material, or surface condition of the measurement object.
As shown below, the value can be corrected by setting two points of "a" and
"b" respectively to the point "A" and "B". The point "A" means the actual
measurement value, and the point "a" means the correction value. The point
"B" and the correction value "b" can be set as same as the above.
Output
Measurement value A
Correction value a
Correction value b
Measurement value A
Before setting the calibration, be sure that the timing input, offset, and zero
set is set to "OFF", the analysis mode is set to "NORMAL", OUT1 is set to
"A" and OUT2 is set to "B" for the output selection. Then, perform the
calibration setting in the appropriate measurement mode.
● How to use
1
Display the "Calibration" screen.
2
Touch the
Change
The screen is changed.
Touch each key to change the screen.
Key
---
Meas A
Measurement value A input screen
Corr a
---
Correction value a input screen
Meas B
---
Measurement value B input screen
Corr b
---
Correction value b input screen
---
Load (top)
Measurement value A load screen
Load (bottom)
Measurement value B load screen
b
: Measurement value
: Correct value
B
Setting screen of sensor head A
key.
Screen
4-19
Chapter 4 Explanation of Functions
Measurement value
A
Correction value
a
Measurement range
Setting screen of sensor head B
For inputting the measurement
value or correction value
For loading the measurement
value from the current value
4