The following sections offer some examples as to how to edit the configuration parameters and
execute Methods.
16.2.10

Procedure to Enter the Transmitter Tag

1. From the My Device menu, make the following menu selections:
Shortcuts > Device Setup > Basic Setup > Device Information > Tag.
2. Click Edit. The Tag screen will be displayed.
3. Key in the tag name (for example: ST 700) which can be a maximum of eight characters.
4. Click OK. The Send to Device screen will be displayed.
5. Click the Tag check box.
6. Click Send to download the change to the Transmitter, or Click Return to continue making
changes.
16.2.11
Selecting the Process Variable (PV) Unit of Pressure
Measurement
Engineering units affect the values of the LRV and URV. After changing the PV
engineering units to the Transmitter, verify changes to the units paramter, the LRV, and the
URV.
The pressure measurement can be displayed in one of the pre-programmed engineering units.
1. From My Device menu, navigate to:
2. Click Edit. You will be warned that if you change the value of the variable it will change the
loop current, which may upset the control process.
Click Yes to continue. The PV Unit screen will be displayed with a list of measurement units,
3.
as follows
:
4. Click the desired PV Unit, and click OK. A Post Edit action message will be displayed,
indicating if you click this value, the variables that use it as the units code will start in the
previous units until this value is sent to the Transmitter.
5. Click OK to continue or Abort to discard the change.
6. Click Send. The Send to Device screen will be displayed.
7. Click the PV Unit check box.
8. Click Send to download the change to the Transmitter or Return to continue making
changes.
Page 140
Shortcuts > Device Setup > Basic Setup > PV Units
inH
O
psi
2
inHg
bar
ftH
O
mbar
2
2
mmH
O
g/cm
2
2
mmHg
kg/cm
ST 700 Series HART/DE Option User's Manual
o
Pa
inH
O@4
C
2
o
kPa
mmH
O@4
C
2
o
Torr
cmH
O@4
C
2
o
Atm
mH
O@4
C
2
MPa