ELECTRICAL CONNECTIONS
3

3.3 Virtual reference for TWM 9000 (C, W and F version)

The virtual reference option on the TWM 9000 flow converter provides complete isolation of the
measurement circuit.
Benefits of virtual reference:
• Grounding rings or grounding electrodes can be omitted.
• Safety increases by reducing the number of potential leakage points.
• The installation of the flowmeters is much easier.
Figure 3-3: Virtual reference
Minimum requirements:
• Size: ≥ DN10
• Electrical conductivity: ≥ 200 µS/cm
• Electrode cable: max. 50 m / 164 ft, type DS

3.4 Connection diagrams

INFORMATION!
For the connection diagrams please refer to the documentation of the applicable signal
converter.