Table of Contents

Advertisement

Operating Manual
Multi-Purpose Detector (MPD)

Advertisement

Table of Contents
loading

Summary of Contents for Honeywell MPD Series

  • Page 1 Operating Manual Multi-Purpose Detector (MPD)
  • Page 3: Table Of Contents

    Table of Contents Safety5 1 Introduction 8 1.1.1 Naming Conventions ............8 12 Principle of Operation 9 1.2.1 Catalytic Bead Sensor MPD-**-CB1 ........9 13 Accessories 9 1.3.1 Collecting Cone .............10 1.3.2 Weatherproof Cap ............10 1.3.3 Sample / Calibration Gas Flow Adaptor ......11 1.3.4 Gassing Point Assembly ..........11 1.3.5 Duct Mount Kit ...............12 1.3.6 Filters ................12...
  • Page 4 81 MPD Catalytic Bead and IR Replacement Sensor Cartridges 25 82 Terminal Housings 25 83 Maintenance Spares 25 84 MPD Calibration Gases26 9 Specifications 26 91 General Specifications 26 92 Performance Data 26 93 MPD-CB1 28 9.2.1 MPD-CB1 Cross Sensitivity ...........28 94 MPD-IV129 95 MPD-IF1 30 96 MPD-IC1 31...
  • Page 5: Safety5

    Read and understand this manual completely before operating or servicing. If any further details are required which do not appear in this manual contact Honeywell Analytics or their agent. The Code of Practice regarding ‘Selection, Installation and Maintenance of Electrical »...
  • Page 6 CAUTIONS To maintain safety standards, regular maintenance, calibration and operation of the » equipment by qualified personnel is essential. Do not modify or alter the sensor construction as essential safety requirements » may be invalidated. When replacing sensors, dispose of used sensors in accordance with local »...
  • Page 7 The user of this manual should ensure that it is appropriate in all details to the exact equipment to be installed and/or operated. If in doubt, the user should contact Honeywell Analytics for advice.
  • Page 8: Introduction 8

    Reference drawing number 1226E0351 for specific installation requirements. Honeywell recommends periodic bump testing of the sensor to ensure proper operation and compliance with the functional safety rating of the installation. 111 Naming Conventions MPD sensor models are in the format MPD-XX-YYY.
  • Page 9: 12 Principle Of Operation 9

    3/4” NPT or M25 Threads Sensor Cell Sensor Retainer and Locking Screw 2.080 3.016 [76.62] [52.83] 1.875 [47.63] 20AWG Stranded Wire 7.5 [190.5] Nominal Length 2.016 (All dimensions reference) [51.21] Figure 1. Multi-Purpose Detector (MPD) 12 Principle of Operation 121 Catalytic Bead Sensor MPD-**-CB1 The CB1 sensor, catalytic bead type contains two elements which are heated by a defined constant electrical current from the transmitter.
  • Page 10 131 Collecting Cone (02000-A-1642) The detection of a lighter than air gas is enhanced by the use of a Collecting Cone. The cone fits onto the Sensor accessory thread in place of the Filter Housing and retains the filter. The appropriate filter must be fitted. A nozzle on the cone permits gassing of the Sensor with the cone in position.
  • Page 11 133 Sample / Calibration Gas Flow Adaptor (1226A0411) A Flow Adaptor provides a facility to allow sampling of a closed system by means of two pipelines. The Flow Adaptor is fitted to the sensor accessory thread by a locking ring that enables the Flow Adaptor to be removed without disconnecting the pipeline.
  • Page 12 135 Duct Mount Kit The duct mounting kit (S3KDMK), combined with the MPD Interface Adaptor (1226A0382), can accommodate the MPD to detect flammable gases in a duct application. The duct mount kit includes the adaptor, gasket and required fasteners. The MPD Interface Adaptor includes only the adaptor and requires the S3KDMKduct mount kit.
  • Page 13: Installation 13

    2 Installation WARNING The Code of Practice regarding ‘Selection, Installation and Maintenance of Electrical Apparatus for use in Potentially Explosive Atmospheres’ must be complied with at all times. Refer to the appropriate local and/or national regulations relative to the installation site. Elsewhere the appropriate local and/or national regulations should be used.
  • Page 14: 24 Wiring 14

    24 Wiring NOTe: To maintain EMC integrity, wiring must be shielded by either an integral shield or run through conduit or pipe. Shield should provide 90% coverage. The MPD Sensor must be operated from a transmitter supplied by a Class 2 Power Supply.
  • Page 15: 25 Fitting Sensors 15

    25 Fitting Sensors WARNING Before installing the sensors, isolate the power supply by disconnecting or removing the associated control module from the installation. CAUTION • The sensitivity of catalytic sensors is impaired by silicone compounds. Do not expose the sensors to silicones or silicone-based products. •...
  • Page 16: Fitting Accessories 16

    3 Fitting Accessories 31 Collecting Cone (02000-A-1462) To fit the Collecting Cone to a sensor, carry out the following: 1. Remove the Filter Housing and gasket from the sensor. 2. Fit the Stainless Steel Filter. 3. Screw the Collecting Cone onto the sensor accessory thread and tighten firmly by hand.
  • Page 17: Maintenance17

    4 Maintenance 41 General Maintenance consists of cleaning the sensor and accessories, replacing gasket and the Hydrophobic Barrier and gassing the sensor when testing the system. CAUTIONS • The sensitivity of Catalytic Sensors is impaired by silicone compounds. Do not expose to silicones or silicone based products. •...
  • Page 18: Faults 18

    Internal O-ring (critical to performance times) Cell Alignment Slot Sensor Cell IR or Cat Bead Sensor Retainer and Locking Screw Figure 9. Replacing Plug In Sensor 5 Faults The following table provides a list of possible faults related to the sensor together with possible causes and remedies.
  • Page 19: Functional Gas Test (Bump Test) 19

    6 Functional Gas Test (Bump Test) 61 Sensor without Accessories Where there are no accessories fitted, it is recommended that a Flow Adaptor is used when gassing the sensor. Where this is not possible, a suitable plastic bag may be used. 62 Sensor with Collecting Cone or Weatherproof Cap 1.
  • Page 20: Calibration 20

    The calibration procedure should only be performed by qualified personnel. NOTe: Honeywell Analytics recommends that the maximum calibration interval be 6 months or in accordance with customer site procedures, whichever is sooner to assure the highest level of safety. Correct operation of each sensor/detector should be confirmed with test gas of known concentration before each use.
  • Page 21 Figure 11. MPD with Flow Adaptor NOTe The Gas Calibration menu is for both Zero and Span Calibration. 3. Connect the Flow Adaptor (using either gas pipe) to the regulated cylinder containing a known concentration of the target gas at approximately the sensor alarm point (e.g.
  • Page 22 72 Cross Calibration procedure for MPD-CB1 CAUTION Where the user calibrates any sensor using a different gas, responsibility for identifying and recording calibration rests with the user. Refer to local regulations where appropriate. When the MPD-CB1 Combustible LEL sensor is to be calibrated with a gas which is different to the gas or vapor to be detected, the following cross calibration procedure should be followed: NOTe...
  • Page 23 * Rating of * Rating of Gas to be Detected Calibration Table 2. Meter Settings 4. If a sensor is to be used to detect a gas other than that for which it was calibrated, the required correction factor may be obtained from Table 3. The meter reading should be multiplied by this number in order to obtain the true gas concentration.
  • Page 24: 73 Mpd-Cb1 Flammable Sensor Operational Life 24

    NOTe Honeywell Analytics recommends that users verify the accuracy of their instruments using test gases wherever possible. Cross-referred measurements should be used as a guide only, not as absolute values.
  • Page 25: Replacement Parts 25

    8 Replacement Parts 81 MPD Catalytic Bead and IR Replacement Sensor Cartridges Sensor Cartridge Maximum Selectable Default Target Gas Increment Type Part No Range Range Range Carbon MPD-IC1 1226-0301 5.00 %Vol 1.00 to 5.00 %Vol 1.00 %Vol 5.00 %Vol Dioxide 5.00 %Vol or 1.00 to 5.00 %Vol MPD-IV1...
  • Page 26: 84 Mpd Calibration Gases26

    84 MPD Calibration Gases Sensor Type Target Gas Cal Gas Range Cal Gas P/N Cal Gas Description MPD-IC1 Carbon Dioxide 1.50 to 3.5 %Vol Contact HA 2.5 %VOL CO in air MPD-IV1 Methane 1.50 to 3.5 %Vol GFV352 2.5 %VOL CH in air or LEL MPD-IF1 Flammables...
  • Page 27 Operating Temperature Sensor Drift Over Operating Operating Type Time Pressure Air Speed MPD-IC1 < 3%/yr -20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec MPD-IV1 < 3%/yr -20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec MPD-IF1 <...
  • Page 28: 93 Mpd-Cb1 28

    93 MPD-CB1 Characteristic Description Gases Detected: Methane gas in the %LEL range. Combustible gases in the %LEL range with sensitivity dependent on gas type. Measuring Range: 0 – 100%LEL Methane Other gas ranges may vary. Environmental - XNX MPD -40°C to +65°C/-40°F to +149°F Sensor Storage Temperature Operating Temperature Range: -40°C to +65°C Operating Humidity Range: Continuous: 20 to 90% RH non-condensing...
  • Page 29 94 MPD-IV1 Characteristic Description Gases Detected: Methane gas in the %Vol and %LEL range. Measuring Range: 0 – 5.00%Vol 0 – 100%LEL Operating Temperature Range: -20°C to +50°C Operating Humidity Range: 0 to 95% RH non-condensing Operating Pressure Range: 80 to 110 kPa Warm Up Time: 3 minutes Response Time: T90 = <30 seconds Time to reach 90% of a step change in gas concentration of...
  • Page 30 95 MPD-IF1 Characteristic Description Gases Detected: Propane gas in the %LEL range. Combustible gases in the %LEL range with sensitivity dependent on gas type. Measuring Range: 0 – 100%LEL Propane Other gas ranges may vary. Operating Temperature Range: -20°C to +50°C Operating Humidity Range: 0 to 95% RH non-condensing Operating Pressure Range: 80 to 110 kPa Warm Up Time: 3 minutes...
  • Page 31 96 MPD-IC1 Characteristic Description Gases Detected: Carbon Dioxide in the %Vol range. Measuring Range: 0 – 5.00%Vol Operating Temperature Range: -20°C to +50°C Operating Humidity Range: 5 to 95% RH non-condensing Operating Pressure Range: 80 to 110 kPa Warm Up Time: 3 minutes Response Time: T50 = <30 seconds, T90 = <70 seconds with SPXCDWP, T50 = <60 seconds, T90 = <120 seconds Time to reach 50% and 90% of a step change in gas...
  • Page 32 9.7 MPD Certifications by Part Number Series UL/CSA Certifications by Part Number Series Hazardous Location - UL UL 1203 Hazardous Location - CSA CAN/CSA C22.2 No. 30 M-1986 N/A N/A N/A N/A N/A N/A Performance with XNX Transmitter XNXUT*V-***** CSA C22.2 No. 152 N/A N/A N/A N/A N/A N/A N/A Performance with XNX Transmitter XNXUT*V-*****, XNXBT*V-***** FM 6310 / 6320...
  • Page 33 971 MPD Hazardous Area Approvals Description Class I, Division 1 Groups B,C,D Class I Division 1 l l l l l l l l Groups B,C, D Class II Division 1 l l l l l l l l Groups F & G AEx d IIB+ H2 T4 -40 °C ≤Tamb≤...
  • Page 34: 94 Mpd-Iv129

    972 MPD Certification Labels Refer to the table below for label details. UL/CSA/FM Certification Label for MPD-UTCB1 Serial Number Sensor Type Model Number As to Fire and Explosion Only Enclosure for use in Hazardous Areas CAUTION: To Prevent Ignition of Hazardous Atmospheres disconnect power before servicing 39YC FM Approved to Class I, Zone 1,...
  • Page 35: Warranty 34

    Analytics Service Representative for information on Service Contracts Warranty Conditions 1. The Honeywell Analytics (HA) Limited Product Warranty only extends to the sale of new and unused products to the original buyer where purchased from HA or from a HA authorized distributor, dealer or representative. Not covered are: consumable items such as dry-cell batteries, filters and fuses or routine replacement parts due to the normal wear and tear of the product;...
  • Page 36 If you purchased your HA product as a consumer, the above warranty conditions do not affect your rights under any applicable consumer protection legislation. Honeywell Analytics reserves the right to change this policy at any time. Contact Honeywell Analytics for the most current warranty information.
  • Page 37: Ec Declaration Of Conformity 36

    11 EC Declaration of Conformity EC Declaration of Conformity The undersigned of For and on behalf of the importer l l I Switzerland United States Declares that the products listed below MPD Adaptor The MPD (Multi Purpose Detector) is a serviceable flameproof and dust ignition protected housing, offered with either Catalytic Bead or Infrared sensors for the measurement of flammable and toxic gasses Are in conformity with the provisions of the following European Directive(s), when installed, operated, serviced and maintained in accordance with the installation/operating instructions contained in the product documentation:...
  • Page 40 Lincolnshire, IL 60069 Tel: +1 847 955 8200 Toll free: +1 800 538 0363 Fax: +1 847 955 8210 detectgas@honeywell.com Asia Pacific Honeywell Analytics Asia Pacific #508, Kolon Science Valley (I) 187-10 Guro-Dong, Guro-Gu Seoul, 152-050 Korea Tel: +82 (0)2 6909 0300 Fax: +82 (0)2 2025 0329 analytics.ap@honeywell.com...

This manual is also suitable for:

Mpd-if1Mpd-cb1Mpd-ic1Mpd-iv1

Table of Contents