Troubleshooting:

If communication cannot be established, the monitor will display one of the following error messages; Error Code E203, E204 or E205,
Scale Disconnected:
E203 error code:
The Bluetooth Adapter is not plugged in properly to the back of the monitor.
Solution:
Check the connection on the rear panel and retry.
E204 error code:
Bluetooth Adapter is not paired to the scale.
Solution:
Check the Bluetooth Adapter. Make sure that the Bluetooth Adapter is powered ON and that it is properly connected to the
Adapter Cable, and that the Adapter Cable is properly connected to the DM Monitor at the Scale port on the rear panel of the DM
Monitor. Retry.
P4607EN.01
ERROR CODE: E203
Scale Disconnected
ERROR CODE: E204
Scale Disconnected