E205 error code:
No Bluetooth connection because the scale is not powered, the scale batteries are absent, improperly installed, or
need replacement, or the scale is too far away from the monitor.
Solution:
To correct the error, check that there is power to the scale, or that it is located within range of the monitor (approximately
10 meters, or 30 feet). Reschedule another vitals collection to retry.
ERROR CODE: E205
Scale Disconnected
Changing batteries (if applicable)
To change the batteries, turn the scale over. Use a Phillips screwdriver to remove the screw retaining the battery access panel to the scale.
Observe the orientation of the four AA batteries inside the scale. Remove the batteries and replace with four new AA batteries for best
results.
NOTE: The DM Monitor MUST BE RESTARTED after changing the batteries. This is because the Bluetooth wireless scale is not
configured to give a zero weight reading ("zeroed") on power up. The scale is only zeroed during monitor startup when the monitor does
the scale self test. This is unique to the Bluetooth wireless scale – including when the AC adapter is being used. For other scales, the DM
Monitor will automatically zero the scale on scale power up.
Zeroing the scale
After changing batteries on the Bluetooth wireless scale, return the scale to its normal position and assure that there are no items on the
scale. Re-start the DM Monitor. The monitor will conduct a scale self-test and zero the scale for optimal accuracy of weight readings. Check
the zeroing to assure that zero "0" appears on the monitor display.