4.6 Error Messages
Message
ROM ERROR
PRESS "CL" KEY TO CLEAR
MESSAGE
RAM ERROR
PRESS "CL" KEY TO CLEAR
MESSAGE
PARAMETER ERROR
PARAMETERS ARE SET TO
DEFAULT VALUE.
LOG INFORMATION ERROR
LOG INFORMATION IS
INITIALIZED.
EEPROM ERROR
INSTRUMENT INFO. IS
INITIALIZED.
WL DRIVE MECHANISM
ERROR
WL MOTOR DRIVE DRIVER
ERROR
WL CALIBRATION ERROR
PRESS "CL" KEY TO CLEAR
MESSAGE
THERMO CELL ERROR
THERMO CELL IS NOT
CONNECTED
PRESS "CL" KEY TO CLEAR
MESSAGE
Description
ROM is faulty.
RAM is faulty.
Parameter is wrong.
Log information is wrong.
Instrument information is
wrong.
Initialize test is
unsuccessful.
Motor driver for
wavelength is faulty.
The 656 nm bright line of
D
lamp is not found in
2
self-diagnosis (at
initialization).
Thermostatic flow cell is
faulty.
Thermostatic flow cell
cannot be connected.
4 - 21
Remedy
Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
1) Turn on power supply again.
2) Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
1) Turn on power supply again.
2) Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
1) Turn on power supply again.
2) Contact the nearest maintenance
service company authorized by
Hitachi High-Technologies.
1) Clear the message by CL key.
2) Confirm proper mounting of the
flow cell, and then turn on power
supply again.
3) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
1) Turn on power supply again.
2) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
1) Clear the message by CL key.
2) Confirm proper mounting of the
flow cell, and then turn on power
supply again.
3) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
1) Confirm proper mounting of the
thermostatic flow cell, and then
turn on power supply again.
2) Contact the maintenance service
company authorized by Hitachi
High-Technologies.
1) Clear the message by CL key.
2) Confirm proper mounting of the
thermostatic flow cell, and then
turn on power supply again.
3) Contact the maintenance service
company authorized by Hitachi
High-Technologies.