(g) SPURGE
Syringe purging can be performed. The wash fluid a
syringe aspirated is discharged to the port4 side of the
syringe valve.
STEP RACK VIAL VOL(µL) INJ
1
FUNC
FUNC (TUBE=1, MOVE=2, HEIGHT=3, SYR=4,
7
VALVE=5, SVALVE=6, SPURGE=7, WPUMP=8)
ESC
7 ENT
SYRINGE PURGE
<1-20>
3 ENT
ESC
SYRINGE PURGE
<1-5>
5 ENT
ESC
MOVING
IDLE WAIT!
PRESS FUNC KEY OR IDLE KEY
MOVING
IDLE WAIT!
Prior to pump washing, be sure to disconnect the stop
NOTE:
valve from port 7 of the syringe valve.
1
1
10.0
(STROKES SPEED)
3
(STROKES SPEED)
3
IDLE
Monitor screen reappears.
3 - 49