No.
Error Message
13
INJ TIMING ERROR
14
NO TUBE
(Note: In execution of
a sequential program,
it is terminated
without indication of
"NO TUBE" on the UI
pad. In execution of
the TUBE command,
"NO TUBE" is
indicated on the UI
pad. Refer to
Section 3.3.2.)
15
STOPPED BY
EXTERNAL ERROR
16
SEQ PROGRAM
BUFFER
OVERFLOW
17
ERROR-X
18
ERROR-Y
19
ERROR-Z
Table 4-3 Autosampler Main Unit Errors (continued)
Possible Cause
1. Any pump other than the
Chromaster 5110 pump is
combined for use with the
autosampler.
2. The flow rate specified for
the pump is too low.
1. No sample vials are set.
2. The clearance between
the top face of sample vial
and the detection lever is
inappropriate. Or the
detector lever is caught.
3. The electric system
malfunctions.
An error has occurred in other
unit connected.
Communication is abnormal.
1. The sample rack or
sample vial is blocking
movement.
2. There is an obstacle on
the moving route of arm X
(in left-right direction).
3. The electric system
malfunctions.
1. There is an obstacle on
the moving route of arm Y
(in front-rear direction).
2. The electric system
malfunctions.
1. There is an obstacle on
the moving route of needle
Z (vertically).
2. The electric system
malfunctions.
4 - 20
Remedy
1. Set the Injection Timing
function to "NO USE".
2. Adjust the minimum flow rate
properly. Alternatively, set
the Injection Timing function to
"NO USE".
1. Set sample vials.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
Eliminate the cause of the error to
release the error status.
Contact your nearest service
office of Hitachi High-
Technologies Corporation sales
representative.
1. Set the sample rack or sample
vial correctly.
2. Remove the obstacle.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. Remove the obstacle.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. Remove the obstacle.
2. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.