HP 8340B Manual page 59

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

4-12.
PULSE MODULATION RISE AND FALL TIME TEST
NOTE
THIS
TEST
IS
NOT APPLICABLE TO INSTRUMENTS EQUIPPED WITH
OPTION
006, DELETE PULSE
MODULATION.
Specification
Table 4-25.
Pulse
Modulation
Rise and
Fall Time Test
Specification
For
CW
mode and frequencies
2:400 MHz
only:
Rise (TF)) and
Fall
(Tp) Times:
<25
nanoseconds
Description
The DUT's RF output frequency
is
down converted to 50
MHz
using
a
mixer and
a
local oscillator.
The
50
MHz
IF signal
is
amplified and applied to an oscilloscope.
The DUT
is
pulsed
using
a
pulse
generator. The pulse generator output
is
also applied to the oscilloscope. The oscilloscope
is
used to
measure the
pulse
envelope
rise and fall times.
Refer
to Figure
4-12
Pulse Definitions.
RF
PULSE
OUT
INPUT
PULSE
-J
TBF-RF
TV
-Inpu
Tn
-Dela
Vp
-RFF
1
90% ——
10%
50%——
100%——
———TO——
/
-————Ty-
Pulse
Length
t
Pulse Lengt
yTime
'ulse
Amplitu
^R
r
/
\-
'
•RF
-'
h
de
VQR
T
°R
^
'
1
-
^
J\
N
Tp
-RF
Pulse
Ri»
Time
TF
-RF
Pulse Fall
Time
VQn-Overshoot and
Ringing
Vp
-Video Feedthrough
:
4
iVp
--I-*
^r——
VF
/i
Figure 4-12.
Pulse Definitions
HP 8340B/41B
Performance Tests

Advertisement

Table of Contents
loading

This manual is also suitable for:

8341b

Table of Contents