4-6.
EXTERNAL LEVELING (Cont'd)
Equipment
Digital
Voltmeter (DVM)
..........................
Crystal
Detector
................................
Capacitor (1000 pF)
.............................
Adapters
APC 3.5 (f) to APC 3.5 (f)
.......................
(1 required
if
DUT
is
HP 8340B)
TYPE
N
(m) to APC 3.5 (f)
.....................
(1
required
if
DUT
is
HP 8341 B)
BNC (f)
to Dual Banana
.......................
BNC
Tee
(m) (f) (f)
............................
Attenuators (required for Option
001
or Option 005 only)
10 dB
.......................................
20 dB
........
HP
3456A
........
HP 8473C
.
HP P/N 0160-4574
.
HP P/N
5061-5311
.
HP P/N 1250-1744
.
HP P/N 1251-2277
.
HP P/N 1250-0781
HP
8493C
Opt. 010
HP
8493C
Opt.
020
Procedure
1.
Turn DUTs LINE switch ON. Press
[INSTR PRESET] [CW] [4]
[.] [5]
[GHz] [POWER LEVEL] [0]
[dB(m)].
2.
Connect equipment
as shown
in
Figure
4-6
(note 1000
pF capacitor shunting DVM
input). Allow
the equipment to warm up for one
hour.
NOTE
In
step
3,
XTAL mode
is
selected.
In
this
mode the [dB(m)] terminator
key selects dBV. The dBV value selected corresponds
to
a
voltage
at
which the
DUTs
EXT INPUT
will
be leveled (see
Table
4-14).
The
selected dBV
is
displayed
in
the
ENTRY DISPLAY.
NOTE
If
a
positive polarity crystal detector
is
being
used for
this test,
the
volt¬
ages given
in
Tables 4-12, 4-13,
4-14, and 4-15
will all
be positive.
3.
Press
[XTAL] [6] [dB(m)]. DVM
reading should be within the values shown on the
first line
in
Table
4-12.
Table 4-12.
External
Leveling
Limits
(With No Attenuator)
dBV
Selected
+6
0
-10
-20
DVM Indication (V)
Lower Limit
-1.883
-0.944
-0.2983
-0.0942
Ideal
-1.995
-1
000
-03162
-001000
Upper Limit
-2.114
-1.059
-03352
-0.1061
Performance Tests
HP 8340B/41B