5-14.
UNLEVELED RF OUTPUT ADJUSTMENTS (Cont'd)
33. Press the following key sequence
on the DUT:
For an HP 8340B
[INSTR PRESET]
[START FREQ] [1] [9] [.] [8] [GHz]
[XTAL]
[SWEEPTIME] [2] [0] [0] [msec]
[SAVE]
[1]
[SWEEPTIME] [AUTO]
[SAVE] [2]
[SINGLE]
[SAVE] [3]
[RECALL] [2]
For an HP 8341
B
[INSTR PRESET]
[START
FREQ] [1] [3] [.] [3] [5]
[GHz]
[XTAL]
[SWEEPTIME] [2] [0] [0] [msec]
(500 mS
if
the
DUT
has Option 003)
[SAVE] [1]
[SWEEPTIME] [AUTO]
[SAVE] [2]
[SINGLE]
[SAVE] [3]
[RECALL] [2]
34. Adjust
the
scope's
vertical and horizontal sensitivity to view the
swept display. The horizontal
gain
should be set so the trace
covers
the
full
display from left to
right.
35. Access Calibration Constant 75 by pressing the following key sequence on the DUT:
[SHIFT]
[GHz] [7] [5] [Hz]
[SHIFT] [MHz] [1] [2] [Hz]
[SHIFT] [kHz] [2] [2] [Hz]
NOTE
The
range for the delay compensation Cal constants, numbers
2
through
8, is
0
to
131. If
the
Cal
Constants do not
provide sufficient
range
for
the
delay compensation,
Cal Constant
8
should be lowered (Cal Constant
7
for
the
HP
8341B), A28R22 (DYO)
and A28R19 (DYS) readjusted, and then
the SYTM DELAY
COMPENSATION ADJUSTMENTS
BY BAND procedure
should
be repeated.
36. Adjust
A28R22
(DYO) for optimum power
in
the first half of the display. Ignore the small
power
dropout
at
the
beginning of the
sweep,
this will
be eliminated later.
37. Adjust
A28R19 (DYS)
for
optimum power
in
the second half of the display. Iterate
between
steps
36
and 37
until
optimum
power
is
achieved across the band.
38. Using the DUT's rotary knob, adjust Cal Constant
75 (range
is 0
to 500)
to optimize the power
in
the
first
division (left hand side) of the display (i.e., eliminate the small
power
dropout).
HP 8340B/41B
Adjustments
5-81