4-1.
INTERNAL TIME BASE AGING RATE (Cont'd)
OSCILLOSCOPE
FREQUENCY STANDARD
^
son
FEEDTHROUQH
TERMINATION*
•T
OUTPUT
•REQUIRED
ONLY
FOR
OSCILLOSCOPES
NOT
HAVING
50
OHMS
INPUT
CAPABILITY
Figure 4-1.
Internal Time
Base Aging Rate
Test
Setup
NOTE
Be sure the
HP
8340B/41B has been connected
to the AC
power
line for
30 days before beginning this test.
If
the instrument
was disconnected
from the ac power line for less than 24 hours,
it
only has to
be connected
to AC
power for 24
hours
before proceeding.
Procedure
1.
Connect the equipment
as shown
in
Figure
4-1-
2.
Adjust the oscilloscope external triggering
controls for
a
stable display of the HP 8340B/41B
10
MHz
REF OUTPUT
signal.
3.
Measure
the
time required for
a
phase change of
360 degrees.
Record the time (T1)
in
seconds.
T1
=
second(s)
Performance Tests