8-7.
Low Output Power Accuracy Test and
RF
Attenuator Calibration
Description
This test:
Adjusts the instrument's attenuator calibration constants to compensate for errors
in
the
90 dB
step
attenuator
(for
those instruments
equipped with
a
90 dB
step attenuator).
Verifies
power accuracy
at each attenuator step.
Sums
the
attenuator performance data
with
output power accuracy data and compares the
results with the instrument's performance specifications.
NOTE:
This test
will
not run unless the
IF
attenuator has been characterized (see section 6) and the
output
power
accuracy test has been run.
The
data
file
generated
for the IF
attenuator must
be on disc
1,
and the data
file
for the output power accuracy test must be
on the
data disc you are
using.
Test Equipment
Local Oscillator
.....................................
HP
8340B
Option
H01
Spectrum Analyzer
...........
...............................
HP
3585A
Amplifier
.....................................................
HP
8447F
Step Attenuator
(Calibrated at
1
MHz)
..............................
HP
355D
Attenuators:
10
dB
..........................................
HP
8493C
Option 010
20 dB
............................................
HP 8493C Option 020
Mixer
.............................................
HP Part No.
0955-0307
Low Pass
Filter
.........
........................
HP Part No. 9135-0260
Adapters:
3.5
mm (f)-(f)1
...........
.........................
HP Part No. 5061-5311
3.5
mm
(0
-
Type-N
(m)2
...........................
HP Part No. 1250-1744
SMA (m)
-
BNC
(f)
...............................
HP Part No. 1250-1200
Cables:
SMA
(m)-(m)
....................................
HP Part
No.
08340-20125
BNC (m)-(m) (7 Required)
............................
HP
Part No. 8120-1840
1
Required for an HP
8340A/B
2
Required
for
an HP 8341A/B or
8340A H02
NOTE:
For
maximum
drive
into
the mixer LO port, connect the mixer directly to the LO (local
oscillator) RF output.
NOTE:
The
test setup for the adjust and measure portions of the RF attenuator calibration
is
critical.
The
tests and adjustments are performed at low RF output power levels (to —90 dBm).
A
defective
cable or
adapter, or
a
loose cable connection
in
the test setup can cause
a
misadjustment of the
instrument or failure of the low output power accuracy test.
If
the
test results
indicate that the instrument
failed low output
power accuracy, check
the test setup and re-run the selected test (either adjust or
measure).
Attenuator Calibration and Operation Verification Test Software
Tests/Utilities
8-9