Connections, Series I And Ii Systems; Side Panel - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Gantry
3.2.6

Connections, Series I and II systems

I, II
3.2.6.1

Side Panel

On Series I and II CardioMD systems, all connections are routed via the side panel, with the
exception of mains power. Power is applied directly to the mains input located on the power
supply at the rear side of the table console. The side panel is also used to connect the acquisition
PC to the gantry.
Collision indicator
Acquisition PC power
IEEE 1394 FireWire
communication interface
to Acquisition PC
Acquisition PC
power connector
Figure 3.16 Connector panel on side of table console (CardioMD Series I and II)
Acquisition PC Power
The acquisition PC must be powered from this 115 V AC outlet. The output is protected by a
circuit breaker.
IEEE 1394 (FireWire)
The IEE 1394 FireWire connector interfaces the acquisition PC to the gantry.
3-20 CardioMD Service Manual
E-Stop
ON indicator
Motor power
indicator
E-Stop indicator
Interface to hand
controller
Trigger input from
ECG gate
Button for resetting
acquisition PC
circuit breaker
Interface to E-Stop
located next to
acquisition PC
1MAN0158-H11 04-2005 CSIP Level 1