27. Mount the spring onto the actuator
bracket.
28. Check the alignment of optocoupler
detecting the position of the shutter
indicator / manual lever.
When the shutter is in the closed position
(shutter indicator / manual lever in the IN
position), the position must be as indicated on
the photo.
29. If required, the position of shutter
indicator / manual lever can be adjusted by
loosening the two Allen screws shown.
30. Finally, operate the manual lever a few
times and check that it is still moving
smoothly.
31. Reconnect the solenoid cable.
12-64 CardioMD Service Manual
Details
Loosen to adjust
1MAN0158-H11 04-2005 CSIP Level 1