Diagnostic Software Utilities
-hv {value}
-pmt {number}
{value}
-wp
-dyncor
{selector}
{value}
5-38 CardioMD Service Manual
Example:
-> VERSION
EDC Board
Hardware version 9EDC0654-B05
Software version 7CTL0079-A00-b3
Compile time
Set high voltage to specified value. If no value is specified, the actual high-
voltage setting is printed. This is useful when troubleshooting techniques
require the detector high voltage to be changed. The alternative method is
to edit the
file on the acquisition PC and load tables.
hvcal.dat
Example:
-> hv
High Voltage = 1200 V
or
-> hv 1200
High Voltage = 1200 V
Set PMT gain value. If number is zero, all PMTs will be set to the same
value. If no value is specified, the current gain setting will be printed. This
is useful when attempting to manually change a specific PMT gain value or
print the current loaded values.
Example:
-> PMT
PMT gain:
1:947
424
7:648
374
13:380
211
19:233
350
Print PMT work points. This is useful to determine if PMT gains are under
control. All 24 WP numbers must be in the range 20 to 150.
Example:
-> WP
PMT Workpoints
1:
70
82
7:
72
66
13:
63
86
19:
71
57
Note. When reading the PMT workpoints, the order of the row number (1,
7, 13, 19) is reversed.
Set the selected DYNCOR value. If no value is specified, the current
values will be printed. Do the printing to see which selector to use. This is
useful to determine if the DYNCOR calibration file is properly loaded.
Dec
1 2000 - 09:51:20
533
402
560
525
334
281
180
346
383
525
377
298
273
233
206
537
83
76
71
65
76
90
72
83
77 109
66
81
77
97
72
76
1MAN0158-H11 04-2005 CSIP Level 1